FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
MDR report key: 8364397
·
Received February 22, 2019
Report
- Report Number
- 3013756811-2019-07662
- Event Type
- Malfunction
- Date Received
- February 22, 2019
- Date of Event
- January 29, 2019
- Report Date
- February 22, 2019
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00853052007264
- PMA / PMN Number
- P180008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PER TANDEM¿S T:SLIM X2 G6 USER GUIDE: ¿CHANGE YOUR CARTRIDGE EVERY 48 TO 72 HOURS AS RECOMMENDED BY YOUR HEALTHCARE PROVIDER.¿ NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT MULTIPLE INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 113-190 MG/DL. REPORTEDLY, THE CARTRIDGE WAS IN USE FOR 3-4 DAYS WITH HUMALOG INSULIN. TANDEM TECHNICAL SUPPORT EDUCATED CUSTOMER REGARDING CARTRIDGE/INSULIN LABELING. CUSTOMER CHANGED THE CARTRIDGE OR SIMPLY CLEARED THE ALARMS AND RESUMED INSULIN DELIVERY TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 157348 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1000096 | 00853052007264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |