FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP

MDR report key: 8364164 · Received February 22, 2019

Report

Report Number
9614033-2019-00111
Event Type
Malfunction
Date Received
February 22, 2019
Date of Event
January 17, 2019
Report Date
March 5, 2019
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: NO SAMPLES OR PHOTOS DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION. FURTHERMORE, THE LOT WAS INSPECTED ACCORDING TO THE CURRENT WORK INSTRUCTION WITH SATISFACTORY RESULTS FOR THE CHARACTERISTICS REQUIRED FOR ITS APPROVAL INCLUDING FUNCTIONAL TESTS. NO NON-CONFORMING PART THAT COULD BE ATTRIBUTABLE TO THE DEFECT REPORTED BY THE CLIENT WAS PRESENTED. NO DEVIATIONS (QUALITY NOTIFICATIONS) WERE REPORTED DURING THE MANUFACTURING PROCESS. CORRECTIVE AND PREVENTATIVE ACTION WAS OPENED TO INVESTIGATE THIS FAILURE MODE. CAPA 400567.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP HAS BEEN LEAKING. THIS OCCURRED ON 4 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

(B)(6). PMA/510(K)#: K151766, K980987. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP HAS BEEN LEAKING. THIS OCCURRED ON 4 SEPARATE OCCASIONS. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157418 BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON DE MEXICO 7319958 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 Other