MEERA EU WITHOUT AUTO DRIVE
Report
- Report Number
- 8010652-2019-00008
- Event Type
- Malfunction
- Date Received
- February 22, 2019
- Date of Event
- January 18, 2019
- Report Date
- February 22, 2019
- Manufacturer
- HOLGER ULLRICH
- Product Code
- FQO
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- NURSE
Narratives
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF THE LEGAL MANUFACTURER OF THE DEVICE MAQUET GMBH, KEHLER STRASSE 31, RASTATT, GERMANY 76437. EXEMPTION # E2018004. GETINGE USA SALES, LLC, 45 BARBOUR POND DRIVE, WAYNE, NJ 07470. CONTACT PERSON: (B)(4). THE FAILURE DESCRIBED BY THE CUSTOMER COULD BE REPRODUCED AT MAQUET. A SUPPLIER COMPLAINT REVEALED A PROBLEM WITH THE OFFSET OF THE PLASTIC SOCKET. IT MAY BE POSSIBLE THAT IN SOME CASES THE PLASTIC SOCKET IS NOT PRESSED DEEP ENOUGH INTO THE HOUSING SO THAT NOT BOTH PINS SECURE THE PLASTIC SOCKET IN THE HOUSING. FURTHER TESTS BY THE SUPPLIER SHOWED THAT WHEN JUST ONE PIN IS FIXED THE CONNECTION MEETS ITS SPECIFICATION. IN THE INSTRUCTIONS FOR USE (IFU) THE CONNECTION TO MAINS SUPPLY AND THE DETACHMENT OF THE MAINS CABLE IS DESCRIBED. FOR DETACHING IT IS STATED: ¿DETACH THE PLUG AT THE MAIN SOCKET. DETACH THE MAINS CABLE FROM THE MOBILE OPERATING TABLE.¿ THIS MEANS THE CUSTOMER FIRST HAS TO DISCONNECT THE CABLE FROM THE MAINS SOCKED (FROM THE WALL) AND THEN PULL THE PLUG OUT OF THE TABLE. JUST A COMBINATION OF NOT CORRECT ASSEMBLED COLD-DEVICE-PLUG AND NOT FOLLOWING THE IFU CAN LEAD TO THE DESCRIBED FAILURE. THE AFFECTED TABLE WAS REPAIRED BY GETINGE SERVICE TECHNICIAN. GETINGE-MAQUET GMBH PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.
IT WAS REPORTED THAT AT LEAST TWO PEOPLE RECEIVED AN ELECTRIC SHOCK BY TOUCHING A PULLED OUT SOCKET. A SOCKET WAS PULLED OUT OF THE TABLE WHEN THE USER TRIED TO DISCONNECT THE MAINS CABLE FROM THE OPERATING ROOM TABLE. THE MAINS CABLE WAS PLUGGED IN AT THE MAINS SOCKET IN THE WALL. THE CUSTOMER TOLD US THAT THE NURSE TOUCHED TWO PEOPLE WITH THE CABLE. THESE RECEIVED AN ELECTRIC SHOCK. NO MEDICAL INTERVENTION OR SERIOUS INJURY WAS REPORTED. MANUFACTURER REFERENCE # (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154285 | MEERA EU WITHOUT AUTO DRIVE | TABLE, OPERATING-ROOM, AC-POWERED | FQO | HOLGER ULLRICH | 720001B0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |