FDA Adverse Event Malfunction Summary report: N

SYRINGE 20ML LL TIP CONV PAK

MDR report key: 8355951 · Received February 20, 2019

Report

Report Number
9610847-2019-00178
Event Type
Malfunction
Date Received
February 20, 2019
Date of Event
January 31, 2019
Report Date
February 8, 2019
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FMF
UDI-DI
30382903056171
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: (B)(6) SAMPLES WERE RECEIVED FOR EVALUATION. THEY CAME IN THREE PLASTIC TRAYS. THEY WERE VISUALLY INSPECTED UNDER THE MICROSCOPE, FINDING NO FOREIGN MATTER. WHAT WAS OBSERVED IS THE EFFECT CREATED WHEN THE RUBBER STOPPERS TOUCHES THE BOTTOM PART OF THE BARREL, THE STOPPER IS LUBRICATED WITH MEDICAL GRADE SILICONE. AT THE MOMENT IT TOUCHES THE BOTTOM BARREL WALL IT CREATES AN OILY LIKE APPEARANCE. THERE IS NO EXCESS OF SILICONE ON ANY OF THE SAMPLES. SILICONE IS AN INERT, NON-TOXIC MEDICAL SUBSTANCE USED AS A LUBRICANT FOR DISPOSABLE HYPODERMIC PRODUCTS. IT IS AN INTEGRAL PART OF THE SYRINGE, ENABLING IT TO PERFORM AS REQUIRED IN VARIOUS CLINICAL APPLICATIONS AND DOES NOT PRESENT A SAFETY OR EFFICACY ISSUE NOR DOES IT IMPACT PRODUCT FUNCTION. THE SILICONE APPLICATION PROCESS IS DESIGNED TO PROVIDE AN EVEN DISTRIBUTION OF SILICONE ON THE INTERIOR OF THE SYRINGE BARREL, HOWEVER, SOME SILICONE MAY NOT BE PERFECTLY DISTRIBUTED. SILICONE HAS BEEN IN USE IN THIS APPLICATION FOR OVER 20 YEARS, WITH ESTIMATED DISTRIBUTION WELL IN EXCESS OF 25 BILLION UNITS. NO REPORTS ARE KNOWN OF ADVERSE CLINICAL EFFECTS ASSOCIATED WITH THESE PRODUCTS AND UNINTENTIONAL DELIVERY OF SILICONE FLUID LUBRICANT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: THIS IS THE 1ST COMPLAINT FOR THE LOT #S 7173775, 7153545, 7173769 FOR THE SAME DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION OF ISSUES FOR THE COMPLAINT OF BATCH #S 7173775, 7153545, 7173769 DURING THE PRODUCTION RUNS. ROOT CAUSE DESCRIPTION: SILICONE IS AN INERT, NON-TOXIC MEDICAL SUBSTANCE USED AS A LUBRICANT FOR DISPOSABLE HYPODERMIC PRODUCTS. IT IS AN INTEGRAL PART OF THE SYRINGE, ENABLING IT TO PERFORM AS REQUIRED IN VARIOUS CLINICAL APPLICATIONS AND DOES NOT PRESENT A SAFETY OR EFFICACY ISSUE NOR DOES IT IMPACT PRODUCT FUNCTION. THE SILICONE APPLICATION PROCESS IS DESIGNED TO PROVIDE AN EVEN DISTRIBUTION OF SILICONE ON THE INTERIOR OF THE SYRINGE BARREL, HOWEVER, SOME SILICONE MAY NOT BE PERFECTLY DISTRIBUTED. SILICONE HAS BEEN IN USE IN THIS APPLICATION FOR OVER 20 YEARS, WITH ESTIMATED DISTRIBUTION WELL IN EXCESS OF 25 BILLION UNITS. NO REPORTS ARE KNOWN OF ADVERSE CLINICAL EFFECTS ASSOCIATED WITH THESE PRODUCTS AND UNINTENTIONAL DELIVERY OF SILICONE FLUID LUBRICANT. RATIONALE: NA.

Description of Event or Problem · 1

IT WAS REPORTED THAT FIVE SYRINGES 20ML LL TIP CONV PAK EXPERIENCED FOREIGN MATTER CONTAMINATION NOTED PRIOR TO USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149347 SYRINGE 20ML LL TIP CONV PAK PISTON SYRINGE FMF BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 7269869 30382903056171

Patients

Seq Age Sex Outcome Treatment
1 Other