FDA Adverse Event Malfunction Summary report: N

PEGASUS YEL 24GA X 0.75IN QSYTE NON-PVC

MDR report key: 8355773 · Received February 20, 2019

Report

Report Number
8041187-2019-00163
Event Type
Malfunction
Date Received
February 20, 2019
Date of Event
January 30, 2019
Report Date
March 13, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FOZ
UDI-DI
00382903837175
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY REPORT WAS CONDUCTED FOR LOT NUMBER 8170417, AND NO RELATED ABNORMALITIES WERE FOUND. THIS LOT OF PEGASUS 24 GAUGE WAS MANUFACTURED FROM 4/6/2018-4/9/2018, AND THIS IS THE ONLY INSTANCE OF A SAFETY SHIELD ACTIVATION FAILURE OCCURRING IN THIS LOT. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED, WITH NO DEFECTS BEING NOTED DURING FINAL ASSEMBLY OR IN PACKAGING VISUAL INSPECTIONS. YOUR FACILITY WAS ABLE TO SUBMIT SAMPLES FOR EVALUATION; BD ENGINEERS WERE ABLE TO OBSERVE A TILTED V-CLIP DURING THE VISUAL OBSERVATION OF THE DEVICE. THE ROOT CAUSE FOR THIS OCCURRENCE IS THE DESIGN OF THIS COMPONENT. SINCE THE MANUFACTURE OF THIS LOT BD HAS OPTIMIZED THE DESIGN TO PREVENT EVENTS SUCH AS THIS FROM OCCURRING IN THE FUTURE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PEGASUS YEL 24GA X 0.75IN QSYTE NON-PVC HAD A SAFETY SHIELD THAT COULD NOT BE ACTIVATED PROPERLY ACTIVATED AFTER THE CATHETER WAS SUCCESSFULLY PLACED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PEGASUS YEL 24GA X 0.75IN QSYTE NON-PVC HAD A SAFETY SHIELD THAT COULD NOT BE ACTIVATED PROPERLY ACTIVATED AFTER THE CATHETER WAS SUCCESSFULLY PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149499 PEGASUS YEL 24GA X 0.75IN QSYTE NON-PVC INTRAVASCULAR CATHETER FOZ BECTON DICKINSON MEDICAL (SINGAPORE) 8170417 00382903837175

Patients

Seq Age Sex Outcome Treatment
1 Other