FDA Adverse Event Malfunction Summary report: N

SEE H10

MDR report key: 8353812 · Received February 20, 2019

Report

Report Number
2015691-2019-00546
Event Type
Malfunction
Date Received
February 20, 2019
Date of Event
January 29, 2019
Report Date
January 29, 2019
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYE
PMA / PMN Number
SEE H10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CORRECTED DATA: F10, H6. REFERENCE CAPA-20-00141.

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACTURER NARRATIVE: ADDITIONAL INFORMATION WAS RECEIVED AND THE FOLLOWING SECTION(S) WAS UPDATED: SECTIONS ADVERSE EVENT OR PRODUCT PROBLEM, OUTCOMES ATTRIBUTED TO ADVERSE EVENT, EVENT, AND OTHER RELEVANT HISTORY. CORRECTED DATA: UPDATED (ADDITIONAL MANUFACTURER NARRATIVE) THIS MODEL IS NOT SOLD OR MARKETED IN THE U.S. HOWEVER, IT IS SIMILAR TO DEVICE: MODEL #2800; BRAND NAME: CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; PMA #P860057/S001. BIOPROSTHETIC TISSUE VALVES CAN DETERIORATE WITH TIME AND EVENTUALLY FAIL CONTRIBUTING TO REGURGITATION AND/OR STENOSIS. STRUCTURAL VALVE DETERIORATION (SVD) IS THE MOST COMMON REASON FOR BIOPROSTHESIS EXPLANTS/REPLACEMENTS AND ENCOMPASSES MULTIPLE FAILURE MODES, INCLUDING CALCIFICATION, NON-CALCIFIC DEGENERATION, DEHISCENCE, CUSP THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. SUCH FAILURE MODES MAY OCCUR SINGULARLY OR CONCOMITANTLY. IN THIS CASE, CALCIFICATION WAS INDICATED. CALCIFICATION PLAYS A MAJOR ROLE IN THE FAILURE OF BIOPROSTHETIC HEART VALVES. CALCIFICATION OF VALVES OCCURS AS A PROGRESSIVE, TIME-DEPENDENT PROCESS. TISSUE VALVE CALCIFICATION IS INITIATED PRIMARILY WITHIN RESIDUAL CELLS THAT HAVE BEEN DEVITALIZED. INITIAL CALCIFICATION DEPOSITS EVENTUALLY ENLARGE AND GROW INTO A MASS, WHICH STIFFEN AND WEAKEN THE TISSUE AND THEREBY CAUSE THE PROSTHESIS TO MALFUNCTION. THE MINERALIZATION OF A BIOMATERIAL IS GENERALLY ENHANCED AT THE SITES OF INTENSE MECHANICAL DEFORMATIONS GENERATED BY MOTION, SUCH AS THE POINTS OF FLEXION IN HEART VALVES. ULTIMATELY, THE RESULT OF CALCIFICATION IS VALVE FAILURE DUE TO TEARING OR STENOSIS. MANY FACTORS CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION. THESE INCLUDE PATIENT FACTORS (AGE, DISEASE STATE, PHARMACOLOGICAL INTERVENTION, ETC.), MECHANICAL STRESS RELATED TO THE VALVE'S HEMODYNAMIC PERFORMANCE, AND GLUTARALDEHYDE FIXATION OF TISSUE. OF THESE, THE FIXATION PROCESS IS A RELATIVELY MINOR CONTRIBUTOR TO CALCIFICATION FOR EDWARDS' TISSUE VALVES DUE TO ANTI-CALCIFICATION TREATMENTS DURING MANUFACTURING. THOUGH NUMEROUS STUDIES HAVE BEEN CONDUCTED ON PREVENTIVE CALCIFICATION STRATEGIES IN BIOPROSTHETIC HEART VALVES, THE CAUSES OF CALCIFICATION ARE NOT FULLY UNDERSTOOD AND THERE ARE STILL NO MECHANISMS OR MEDICAL THERAPIES WHICH FULLY PREVENT BIOPROSTHESIS FROM CALCIFYING. THE ROOT CAUSE OF THIS EVENT CANNOT BE CONCLUSIVELY DETERMINED WITH THE AVAILABLE INFORMATION. HOWEVER, THE CALCIFICATION OBSERVED IN THIS CASE WAS MOST LIKELY DUE TO A PROGRESSION OF THE PATIENT¿S UNDERLYING VALVULAR DISEASE PATHOLOGY COMBINED WITH THE PATIENT¿S OTHER UNDERLYING RISK FACTORS. THE SUBJECT DEVICE IS NOT AVAILABLE FOR EVALUATION DUE TO CUSTOMER REFUSAL. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS PRIOR TO RELEASE FOR DISTRIBUTION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL REPORTED EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.

Description of Event or Problem · 0

EDWARDS RECEIVED NOTIFICATION FROM THE PATIENT'S SON THAT THIS PATIENT WITH A 19MM PERICARDIAL AORTIC VALVE RE-PRESENTED WITH VALVE CALCIFIC DEGENERATION LEADING TO PERICARDIAL DISCOMFORT ON EFFORT AFTER AN IMPLANT DURATION OF SEVEN (7) YEARS AND SEVEN (7) MONTHS. ECHO REVEALED PEAK GRADIENT OF 113 MMHG AND MEAN GRADIENT OF 67 MMHG. NO AORTIC REGURGITATION PRESENT. A REDO VALVE REPLACEMENT WAS PERFORMED WITH A 19MM NON-EDWARDS AORTIC VALVE AFTER SEVEN (7) YEARS AND 10 MONTHS. THE PATIENT WAS NOTED TO BE IN GOOD CONDITION AFTER THE PROCEDURE.

Additional Manufacturer Narrative · 1

THIS MODEL IS NOT SOLD OR MARKETED IN THE U.S. HOWEVER, IT IS SIMILAR TO DEVICE: MODEL #2800; BRAND NAME: CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS; PMA #P860057/S001. BIOPROSTHETIC TISSUE VALVES CAN DETERIORATE WITH TIME AND EVENTUALLY FAIL CONTRIBUTING TO REGURGITATION AND/OR STENOSIS. STRUCTURAL VALVE DETERIORATION (SVD) IS THE MOST COMMON REASON FOR BIOPROSTHESIS EXPLANTS AND ENCOMPASSES MULTIPLE FAILURE MODES, INCLUDING CALCIFICATION, NON-CALCIFIC DEGENERATION, DEHISCENCE, CUSP THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. SUCH FAILURE MODES MAY OCCUR SINGULARLY OR CONCOMITANTLY. ALTERNATIVELY, NONSTRUCTURAL DYSFUNCTION (NSVD) MAY ALSO PLAY A ROLE IN THE DEVELOPMENT OF VALVULAR REGURGITATION AND/OR STENOSIS. THE ROOT CAUSE OF THIS EVENT CANNOT BE CONCLUSIVELY DETERMINED WITH THE AVAILABLE INFORMATION. HOWEVER, THIS EVENT IS MOST LIKELY IMPACTED BY THE PROGRESSION OF THE PATIENT¿S UNDERLYING VALVULAR DISEASE PATHOLOGY WITH STRUCTURAL VALVE DETERIORATION. THE SUBJECT DEVICE IS NOT AVAILABLE FOR EVALUATION, AS IT REMAINS IMPLANTED IN THE PATIENT. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS PRIOR TO RELEASE FOR DISTRIBUTION. EDWARDS WILL CONTINUE TO REVIEW AND MONITOR ALL EVENTS. TRENDS ARE MONITORED ON A MONTHLY BASIS AND IF ACTION IS REQUIRED, APPROPRIATE INVESTIGATION WILL BE PERFORMED.

Description of Event or Problem · 1

EDWARDS RECEIVED NOTIFICATION FROM THE PATIENT'S SON THAT THIS PATIENT WITH A 19MM PERICARDIAL AORTIC VALVE RE-PRESENTED WITH VALVE DEGENERATION LEADING TO PERICARDIAL DISCOMFORT ON EFFORT AFTER AN IMPLANT DURATION OF SEVEN (7) YEARS AND SEVEN (7) MONTHS. ECHO REVEALED MEAN GRADIENT OF 66MMHG. NOW, DOCTORS ADVISED HER FOR A VALVE REPLACEMENT. INDICATION FOR INITIAL REPLACEMENT WAS SEVERE CALCIFIED AORTIC STENOSIS. NO OTHER DETAILS WERE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149392 SEE H10 REPLACEMENT HEART VALVE DYE EDWARDS LIFESCIENCES 290019MM

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R