FDA Adverse Event Malfunction Summary report: N

PIEZOSURGERY

MDR report key: 8353131 · Received February 20, 2019

Report

Report Number
3003933619-2018-00009
Event Type
Malfunction
Date Received
February 20, 2019
Date of Event
October 15, 2018
Report Date
January 9, 2019
Manufacturer
MECTRON S.P.A.
Product Code
JDX
PMA / PMN Number
K132848
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER MECTRON S.P.A. RECEIVED THE INSERT INVOLVED IN THE EVENT. AN INVESTIGATION WAS CARRY OUT BY THE MANUFACTURER: THE INSERT RETURNED, MT1-10, LOT. NUMBER 18002205, WAS ANALYSED BY AN EXTERNAL LABORATORY. THE RESULT OF THIS ANALYSIS DIDN'T REVEAL ANY CRITICAL PROBLEM ON THE STRUCTURE OF THE INSERT. MOREOVER THE DESIGN, PRODUCTION, IN-PROCESS AND FINAL CONTROLS DOCUMENTS, RELATED TO THE INSERT MT1-10, LOT 18002205 DIDN'T SHOW ANY ANOMALY.

Description of Event or Problem · 1

DURING THE USE IN A MAXILLOFACIAL SURGERY, THE INSERT MT1-10, LOT N. 18002205, EXPIRATION DATE 30 JULY 2023, ETO STERILIZATION BATCH (B)(4), BROKE. NO ADVERSE EVENT OCCURRED TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149382 PIEZOSURGERY MT1-10, INSERT FOR GENERAL BONE SURGERY JDX MECTRON S.P.A. MT1-10 18002205

Patients

Seq Age Sex Outcome Treatment
1 47 YR