MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS
Report
- Report Number
- 1645337-2019-08348
- Event Type
- Injury
- Date Received
- February 19, 2019
- Date of Event
- January 18, 2019
- Report Date
- January 28, 2019
- Manufacturer
- MENTOR TEXAS
- Product Code
- LCJ
- UDI-DI
- 00081317006763
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
ON (B)(6) 2020, UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITH BLUE FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. UPON INITIAL INSPECTION THE DEVICE APPEARS INTACT. LEAK TESTING WAS PERFORMED IN ACCORDANCE WITH MENTOR PROCEDURES AND NO LEAK SITES WERE DETECTED. NO OTHER ANOMALIES WERE DISCOVERED. COMPLAINT WAS NOT CONFIRMED FOR DEFLATION. A MANUFACTURING RECORD EVALUATION (MRE) OF LOT NUMBER 6982497 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON 3/25/2019, IT WAS REPORTED TO MENTOR THAT THE DEVICE EVALUATION AND INVESTIGATION HAVE BEEN COMPLETED. UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITH BLUE FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. UPON INITIAL INSPECTION THE DEVICE APPEARS INTACT. LEAK TESTING WAS PERFORMED IN ACCORDANCE WITH MENTOR PROCEDURES AND NO LEAK SITES WERE DETECTED. NO OTHER ANOMALIES WERE DISCOVERED. COMPLAINT WAS NOT CONFIRMED FOR DEFLATION. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST RECONSTRUCTION PRIMARY WITH MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 800CC AND EXPERIENCED LEFT SIDE DEFLATION OF THE TISSUE EXPANDER. THE PATIENT ALSO EXPERIENCED SEROMA ON THE LEFT BREAST. AS A RESULT, THE PATIENT HAD UNDERGONE REMOVAL AND REPLACEMENT WITH MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 800CC ON (B)(6) 2019. THE PATIENT HAS FULLY RECOVERED AFTER THE SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 145335 | MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS | EXPANDER, SKIN, INFLATABLE | LCJ | MENTOR TEXAS | 3549216 | 6982497 | 00081317006763 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Required Intervention |