FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAG

MDR report key: 8352172 · Received February 19, 2019

Report

Report Number
1920898-2019-00184
Event Type
Malfunction
Date Received
February 19, 2019
Date of Event
January 30, 2019
Report Date
February 22, 2019
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: CUSTOMER RETURNED (8) 3/10CC, 6MM, 31G SYRINGES IN AN OPEN POLY BAG FROM LOT # 6242615. CUSTOMER STATES THAT THE VERY FIRST SCALE MARK IS WAY TOO LOWER THAN IT SHOULD BE. ALL RETURNED SYRINGES WERE TESTED USING THE PLUG GAUGE AND ALL PLACEMENTS OF THE SCALE MARKINGS FELL WITHIN SPECIFICATIONS. NO DEFECTS WERE OBSERVED ON THE SAMPLES. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 6242615. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT 8 SYRINGES 0.3ML 31GA 6MM HALFUNIT 10BAG EXPERIENCED SCALE MARKING ISSUES WITH THE CUSTOMER STATING "THE VERY FIRST SCALE MARK IS WAY TOO LOWER THAN IT SHOULD BE."

Additional Manufacturer Narrative · 1

DEVICE EXPIRATION DATE: UNKNOWN. DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 8 SYRINGES 0.3ML 31GA 6MM HALFUNIT 10BAG EXPERIENCED SCALE MARKING ISSUES WITH THE CUSTOMER STATING "THE VERY FIRST SCALE MARK IS WAY TOO LOWER THAN IT SHOULD BE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143300 SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAG SYRINGE FMF BD MEDICAL - DIABETES CARE 6242615

Patients

Seq Age Sex Outcome Treatment
1 Other