BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP
Report
- Report Number
- 1213809-2019-00225
- Event Type
- Malfunction
- Date Received
- February 19, 2019
- Date of Event
- February 4, 2019
- Report Date
- April 24, 2019
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 30382903096573
- PMA / PMN Number
- K980987
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION SUMMARY: TWO PHOTOS AND 162 SEALED PACKAGED 3ML SYRINGES, CONFIRMED TO BE FROM BATCH #8213796 (P/N 309657) WERE RECEIVED AND VISUALLY EVALUATED. THE PHOTOS DEPICT TWO SYRINGES: ONE SYRINGE WITH A SKEWED SCALE AND PARTIAL MISSING NUMBERS TO THE RIGHT OF THE SCALE, SPECIFICALLY ½, 1 AND PARTS OF 1 ½ AND 2; ONE SYRINGE WITH MISSING PRINT BUT NO SKEW TO THE SCALE ¿ ALL SCALE MARKINGS BETWEEN ZERO LINE AND 1ML WAS MISSING IN THIS SYRINGE. THE 162 RETURNED SYRINGES WERE VISUALLY EVALUATED AND WERE FOUND TO HAVE NO PRINT DEFECTS. ALL HAD SCALES MARKINGS PRINTED CORRECTLY. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. POTENTIAL ROOT CAUSE FOR THE SCALE MARKING DEFECTS IS ASSOCIATED WITH THE MARKING PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 8213795 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS.
IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP SCALE MARKING ARE NOT PRINTED CORRECTLY. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP SCALE MARKING ARE NOT PRINTED CORRECTLY. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143233 | BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 8213795 | 30382903096573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |