FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP

MDR report key: 8350369 · Received February 19, 2019

Report

Report Number
1213809-2019-00225
Event Type
Malfunction
Date Received
February 19, 2019
Date of Event
February 4, 2019
Report Date
April 24, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY: TWO PHOTOS AND 162 SEALED PACKAGED 3ML SYRINGES, CONFIRMED TO BE FROM BATCH #8213796 (P/N 309657) WERE RECEIVED AND VISUALLY EVALUATED. THE PHOTOS DEPICT TWO SYRINGES: ONE SYRINGE WITH A SKEWED SCALE AND PARTIAL MISSING NUMBERS TO THE RIGHT OF THE SCALE, SPECIFICALLY ½, 1 AND PARTS OF 1 ½ AND 2; ONE SYRINGE WITH MISSING PRINT BUT NO SKEW TO THE SCALE ¿ ALL SCALE MARKINGS BETWEEN ZERO LINE AND 1ML WAS MISSING IN THIS SYRINGE. THE 162 RETURNED SYRINGES WERE VISUALLY EVALUATED AND WERE FOUND TO HAVE NO PRINT DEFECTS. ALL HAD SCALES MARKINGS PRINTED CORRECTLY. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. POTENTIAL ROOT CAUSE FOR THE SCALE MARKING DEFECTS IS ASSOCIATED WITH THE MARKING PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. BATCH 8213795 IS CONSIDERED IN COMPLIANCE WITH OUR PRODUCT SPECIFICATION REQUIREMENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP SCALE MARKING ARE NOT PRINTED CORRECTLY. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ TIP SCALE MARKING ARE NOT PRINTED CORRECTLY. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143233 BD LUER-LOK DISPOSABLE SYRINGE WITH BD LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 8213795 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 Other