FDA Adverse Event Injury Summary report: N

PROGAV SYS PED.W/SA 20 A.PRECHAMBER

MDR report key: 8349965 · Received February 19, 2019

Report

Report Number
3004721439-2019-00033
Event Type
Injury
Date Received
February 19, 2019
Date of Event
January 14, 2019
Report Date
October 9, 2019
Manufacturer
B. BRAUN SURGICAL SA
Product Code
JXG
PMA / PMN Number
K103003
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER, REGISTRATION NO. 2916714) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER, REGISTRATION NO. 3003639970). EXEMPTION NUMBER: E2014012 PLEASE PROVIDE THE PATIENT'S AGE, GENDER, WEIGHT, OUTCOME AND MEDICAL INTERVENTION. WHEN ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

INVESTIGATION: VISUAL INSPECTION NO SIGNIFICANT DEFORMATIONS OR DAMAGE WERE DETECTED. PERMEABILITY TEST A PERMEABILITY TEST HAS SHOWN THAT BOTH VALVES ARE PERMEABLE. ADJUSTMENT TEST THE PROGAV VALVE WAS TESTED AND IS ADJUSTABLE TO ALL SPECIFIED PRESSURES. BRAKING FORCE AND BRAKE FUNCTION TEST THE BRAKE FUNCTIONALITY TEST HAS SHOWN THAT THE BRAKE FUNCTION IS FULLY OPERATIONAL AND THE BRAKING FORCE IS WITHIN THE GIVEN TOLERANCES. WE HAVE DISMANTLED THE VALVE. INSIDE BOTH VALVES, WE HAVE FOUND BUILD-UP OF SUBSTANCES (LIKELY PROTEIN). BASED ON OUR INVESTIGATION, WE ARE UNABLE TO SUBSTANTIATE THE CLAIM OF OCCLUSION. THE VALVES OPERATE WITHIN THE SPECIFIED TOLERANCES. HOWEVER, IT IS POSSIBLE THAT THE DEPOSITS OBSERVED INSIDE THE VALVES COULD HAVE TEMPORARILY OCCLUDED THE SYSTEM. AS DESCRIBED IN OUR LITERATURE, THE PROBLEM ENCOUNTERED IS ONE OF THE KNOWN, INEVITABLE RISKS OF HC-THERAPY BY SHUNT IMPLANTS. WE CAN EXCLUDE A DEFECT AT THE TIME OF RELEASE. THE SHUNT SYSTEM MET ALL SPECIFICATIONS OF THE FINAL INSPECTION WHEN RELEASED FROM CHRISTOPH MIETHKE GMBH & CO. KG.

Description of Event or Problem · 0

IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL HAD A BLOCKED VALVE. THE EXPLANTED PRODUCT WERE DELIVERED TO MIETHKE WITH THE RETURN KIT INSIDE AN UNKNOWN LIQUID.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142463 PROGAV SYS PED.W/SA 20 A.PRECHAMBER HYDROCEPHALUS AND ICP MANAGEMENT JXG B. BRAUN SURGICAL SA FV441T 20029222

Patients

Seq Age Sex Outcome Treatment
1 5 YR Other| R