FDA Adverse Event Malfunction Summary report: N

ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE

MDR report key: 8349228 · Received February 19, 2019

Report

Report Number
3001845648-2019-00065
Event Type
Malfunction
Date Received
February 19, 2019
Date of Event
January 16, 2019
Report Date
January 21, 2019
Manufacturer
COOK IRELAND LTD
Product Code
FCG
UDI-DI
00827002315198
PMA / PMN Number
K083330
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

PMA/510(K) # K083330. COOK IRELAND LTD (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI)(IMPORTER). EXEMPTION NUMBER: E2016031. INFORMATION PERTAINING TO SECTION G.1 AS FOLLOWS: IMPORTER SITE CONTACT AND ADDRESS: (B)(4) COOK MEDICAL INCORPORATED (CMI) 1025 ACUFF ROAD P.O BOX 4195 BLOOMINGTON INDIANA 47402-4195. IMPORTER SITE ESTABLISHMENT REGISTRATION NUMBER: 3005580113. THE DEVICE RELATED TO THIS OCCURRENCE UNDERWENT A LABORATORY EVALUATION. A PROXIMAL KINK WAS OBSERVED DURING THE LAB EVALUATION. PRIOR TO DISTRIBUTION, ALL ECHO-25 DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-25 OF LOT NUMBER C1505525 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1505525. THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO; "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE". THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. THE FAILURE OF NEEDLE KINKED/BENT WAS CONCLUDED FROM THE AVAILABLE INFORMATION. A DEFINITIVE ROOT CAUSE FOR THE CUSTOMER COMPLAINT COULD NOT BE DETERMINED AS CIRCUMSTANCES OF USE CANNOT BE REPLICATED IN THE LABORATORY. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO EXCESSIVE FORCE. COMPLAINT IS CONFIRMED AS THE FAILURE WAS VERIFIED IN THE LABORATORY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Description of Event or Problem · 0

IT IS IMPOSSIBLE TO PUT THE STYLET BACK INTO THE NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143399 ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE FCG KIT, NEEDLE, BIOPSY FCG COOK IRELAND LTD G31519 C1505525 00827002315198

Patients

Seq Age Sex Outcome Treatment
1