FDA Adverse Event Malfunction Summary report: N

BD CATHENA¿ BC STRAIGHT CATHETER

MDR report key: 8348629 · Received February 18, 2019

Report

Report Number
8041187-2019-00155
Event Type
Malfunction
Date Received
February 18, 2019
Date of Event
January 25, 2019
Report Date
May 6, 2019
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
LJS
UDI-DI
00382903868049
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ¿ THE RETURNED SAMPLE WAS NOT DECONTAMINATED BY VENDOR 1 SAMPLE WAS RETURNED WITHOUT COVER AND ADAPTER FOR INVESTIGATION IT WAS OBSERVED THAT THE ADAPTER HAD BEEN REMOVED, HOWEVER THE TIP SHIELD IS REMAINED IN THE HUB. SAFETY ACTIVATION FAILURE WAS OBSERVED. THE ADAPTER HAD BEEN REMOVED BUT THE TIP SHIELD IS REMAINED IN THE HUB THE PHOTOS AND ACTUAL SAMPLE RECEIVED SHOW THE REPORTED NON-CONFORMANCE. THE COMPLAINT IS CONFIRMED AND PRODUCT IS OUT OF SPECIFICATION. THE PROBABLE ROOT CAUSE COULD BE DUE THE DAMAGED ADAPTER. HOWEVER, THE ROOT CAUSE COULD NOT BE CONFIRMED AS THE ADAPTER WAS NOT RETURNED FOR INVESTIGATION. CAPA#821876 WAS INITIATED. DHR WAS REVIEWED AND NO SIMILAR QN WAS RAISED ALONG WITH NO ABNORMALITY OBSERVED THAT COULD HAVE INFLUENCED THE ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED WITH THE USE OF THE BD CATHENA¿ BC STRAIGHT CATHETER THERE WAS AN ISSUE WITH SAFETY MECHANISM FAILED TO ENGAGE.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED WITH THE USE OF THE BD CATHENA BC STRAIGHT CATHETER THERE WAS AN ISSUE WITH SAFETY MECHANISM FAILED TO ENGAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139952 BD CATHENA¿ BC STRAIGHT CATHETER INTRAVASCULAR CATHETER LJS BECTON DICKINSON MEDICAL (SINGAPORE) UNKNOWN 00382903868049

Patients

Seq Age Sex Outcome Treatment
1 Other