CLOSED SUCTION SYSTEM FOR ADULTS, 72-HOUR, 12 F, DSE
Report
- Report Number
- 8030647-2019-00024
- Event Type
- Malfunction
- Date Received
- February 15, 2019
- Report Date
- February 19, 2019
- Manufacturer
- AVANOS MEDICAL INC.
- Product Code
- BSY
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
ONE USED SUCTION CATHETER AND ONE USED FLEX CONNECTOR WERE RECEIVED FOR EVALUATION. VISUAL EXAMINATION OF THE DEVICE REVEALED NO DAMAGE. THE CONNECTION BETWEEN THE FLEX CONNECTOR AND DOUBLE SWIVEL ELBOW WAS TESTED AND FOUND TO HAVE A LOOSE CONNECTION. THE COMPLAINT IS CONFIRMED AS REPORTED. ALL INFORMATION REASONABLY KNOWN AS OF 04 MAR 2019 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE HALYARD HEALTH COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORD FOR THE REPORTED LOT NUMBER WAS REVIEWED AND DOCUMENTED THAT THE LOT WAS MANUFACTURED ACCORDING TO THE APPROVED MANUFACTURING PROCEDURES, SPECIFICATIONS, AND THEN RELEASED BY QUALITY ASSURANCE. ALL INFORMATION REASONABLY KNOWN AS OF 15-FEB-2019 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC. AVANOS MEDICAL INC HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL INC COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4).
IT WAS REPORTED THAT THE CONNECTION BETWEEN THE SUCTION CATHETER AND THE FLEX CONNECTOR WAS LOOSE AND DISCONNECTED DURING USE. THE SUCTION CATHETER WAS IMMEDIATELY REPLACED WITH A NEW ONE WITH NO PATIENT INJURY.