FDA Adverse Event Malfunction Summary report: N

ELECSYS 2010 IMMUNOCHEMISTRY ANALYZER

MDR report key: 834368 · Received March 16, 2007

Report

Report Number
1823260-2007-02133
Event Type
Malfunction
Date Received
March 16, 2007
Date of Event
February 11, 2007
Report Date
March 16, 2007
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K961481
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

USER RECEIVED COMPLAINTS FROM PHYSICIANS REGARDING LOW RECOVERY OF B12 ON ELECSYS 2010 WHEN COMPARING THE RESULT TO B12 MEASURED ON THE MODULAR ANALYTICS E170 SYSTEM. THE RECOVERY ON ELECSYS 2010 IS LOWER FOR PT SAMPLES. EIGHT PT SAMPLE RESULTS PROVIDED AS FOLLOWS: PT 1 2010 B12 <22 PMOL/L, E170 89.86 PMOL/L. PT 2 2010 B12 <22.4 PMOL/L, E170 69.25 PMOL/L. PT 3 2010 B12 16.27 PMOL/L, E170 267.5 PMOL/L. PT 4 2010 B12 118.5 PMOL/L, E170 254.3 PMOL/L. PT 5 2010 B12 85.12 PMOL/L, E170 138.5 PMOL/L. PT 6 2010 B12 43.66 PMOL/L, E170 152.8 PMOL/L. PT 7 2010 B12 87.31 PMOL/L, E170 214.4 PMOL/L. PT 8 2010 B12 <22 PMOL/L, E170 99.43 PMOL/L. IF ADDITIONAL INFO IS REC'D, APPROPRIATE NOTIFICATION WILL BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ELECSYS 2010 IMMUNOCHEMISTRY ANALYZER ELECSYS 2010 - JJE JJE ROCHE DIAGNOSTICS 2010 *

Patients

Seq Age Sex Outcome Treatment
1 NA