FDA Adverse Event Injury Summary report: N

DARIO BLOOD GLUCOSE MONITORING SYSTEM

MDR report key: 8338712 · Received February 14, 2019

Report

Report Number
3010606081-2019-00004
Event Type
Injury
Date Received
February 14, 2019
Date of Event
January 17, 2019
Manufacturer
LABSTYLE INNOVATIONS LTD.
Product Code
NBW
PMA / PMN Number
K150817
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

ON (B)(6) 2019, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE READINGS. THE USER REPORTED HE CONSULTED WITH HIS DOCTOR IN ORDER TO CHECK THE READINGS. AT THE DOCTOR, DARIO METER READINGS WERE INCONSISTENT WITH THE LAB RESULTS. THE USER REPORTED HE WAS USING ALCOHOL SWABS BEFORE TAKING MEASUREMENTS. IT IS INSTRUCTED IN DARIO USER GUIDE UNDER 'TESTING INFORMATION AND GUIDELINES': "ALCOHOL CAN AFFECT TEST RESULTS". AT THE MORNING OF (B)(6) 2019, PRIOR TO THE CALL TO DARIO, THE USER DID NOT USE ALCOHOL SWABS AND DARIO BLOOD GLUCOSE READING WAS IN RANGE. AT THE TIME OF THE CASE REPORT, THE USER OPENED A NEW STRIP CARTRIDGE AND THE READINGS WERE IN ORDER. IN ADDITION, THE USER WAS SENT CONTROL SOLUTION. ON (B)(6) 2019, THE USER TESTED WITH THE CONTROL SOLUTION - READINGS WERE IN RANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
132659 DARIO BLOOD GLUCOSE MONITORING SYSTEM NBW NBW LABSTYLE INNOVATIONS LTD. 1021-02 1606211

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization