FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY

MDR report key: 8326122 · Received February 11, 2019

Report

Report Number
1213809-2019-00167
Event Type
Malfunction
Date Received
February 11, 2019
Date of Event
January 24, 2019
Report Date
May 8, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903097037
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE CASE CARTON OF 12 TRAYS WAS RETURNED FOR EVALUATION. THE LABEL ON THE OUTSIDE OF THE CASE CARTON WAS CONFIRMED TO BE FROM BATCH# 8285697. ALL 12 TRAYS INSIDE THE CASE CARTON WERE LABELED AS INCORRECT BATCH# 8280697. THE BATCH NUMBER WAS INCORRECT ON THE TOP WEB AND NO REJECTIONS WERE NOTED DUE TO HUMAN ERROR DURING INSPECTION. QUALITY NOTIFICATION WAS LATER IDENTIFIED WHICH MOST LIKELY CONTRIBUTED TO THE REPORTED CONDITION. ADDITIONAL INSPECTION PERFORMED AND THE BATCH WAS RELEASED BASED ON MEETING OUR RE-QUALIFICATION REQUIREMENT. ROOT CAUSE: SOFTWARE BUG WITH THE VARIABLE PRINTER AND DUE TO HUMAN ERROR DURING INSPECTIONS. PREVENTIVE AND CORRECTIVE ACTION, CAPA#900946, HAVE BEEN INITIATED TO INVESTIGATE THIS REPORTED ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE OUTER LABEL ON THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY STATED THE PRODUCT WAS "LOT 8285697", WHILE THE PACKAGED SYRINGES WERE LABELED AS "LOT 8280697".

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OUTER LABEL ON THE BD LUER-LOK¿ SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY STATED THE PRODUCT WAS "LOT 8285697", WHILE THE PACKAGED SYRINGES WERE LABELED AS "LOT 8280697".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
117988 BD LUER-LOK SYRINGE BULK STERILE PHARMACY CONVENIENCE TRAY PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 8285697 30382903097037

Patients

Seq Age Sex Outcome Treatment
1 Other