BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Report
- Report Number
- 3006948883-2019-00127
- Event Type
- Malfunction
- Date Received
- February 8, 2019
- Date of Event
- January 24, 2019
- Report Date
- March 5, 2019
- Manufacturer
- BD (SUZHOU)
- Product Code
- FOZ
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8233349. OUR RECORDS SHOW THAT THIS IS THE SECOND INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, BASED ON THE SAMPLE PROVIDED OUR QUALITY ENGINEERS WERE ABLE TO OBSERVE A SMALL CRACK IN THE ADAPTER. AFTER REVIEWING THE MANUFACTURING PROCESS THEY DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS A VARIANCE IN THE AUTOLINE'S SWAGE PRESSURE IT IS POSSIBLE FOR THE MACHINE TO APPLY EXCESS FORCE TO THE DEVICE DURING ASSEMBLY, ALLOWING FOR THE RESULTING CRACK TO FORM IN THE DEVICE. CURRENTLY, WE ARE CONDUCTING A LONG TERM STUDY TO DETERMINE THE ROOT CAUSE FOR THIS EVENT, BUT WE ARE ADDRESSING THE ISSUE THROUGH THE IMPLEMENTATION OF MANUAL INSPECTION S FOR CRACKS IN THE ADAPTER HEAD, AND WE ARE FURTHER OPTIMIZING OUR SWAGING PROCESS BY REDUCING DEPTH REQUIREMENTS.
IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD BLOOD LEAKAGE BECAUSE THE CATHETER WAS LIKELY NOT ADHERED TIGHTLY TO THE ADAPTOR. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD BLOOD LEAKAGE BECAUSE THE CATHETER WAS LIKELY NOT ADHERED TIGHTLY TO THE ADAPTOR. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 112216 | BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD (SUZHOU) | 8233349 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |