FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM

MDR report key: 8324399 · Received February 8, 2019

Report

Report Number
3006948883-2019-00127
Event Type
Malfunction
Date Received
February 8, 2019
Date of Event
January 24, 2019
Report Date
March 5, 2019
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8233349. OUR RECORDS SHOW THAT THIS IS THE SECOND INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. ADDITIONALLY, BASED ON THE SAMPLE PROVIDED OUR QUALITY ENGINEERS WERE ABLE TO OBSERVE A SMALL CRACK IN THE ADAPTER. AFTER REVIEWING THE MANUFACTURING PROCESS THEY DETERMINED THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS A VARIANCE IN THE AUTOLINE'S SWAGE PRESSURE IT IS POSSIBLE FOR THE MACHINE TO APPLY EXCESS FORCE TO THE DEVICE DURING ASSEMBLY, ALLOWING FOR THE RESULTING CRACK TO FORM IN THE DEVICE. CURRENTLY, WE ARE CONDUCTING A LONG TERM STUDY TO DETERMINE THE ROOT CAUSE FOR THIS EVENT, BUT WE ARE ADDRESSING THE ISSUE THROUGH THE IMPLEMENTATION OF MANUAL INSPECTION S FOR CRACKS IN THE ADAPTER HEAD, AND WE ARE FURTHER OPTIMIZING OUR SWAGING PROCESS BY REDUCING DEPTH REQUIREMENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD BLOOD LEAKAGE BECAUSE THE CATHETER WAS LIKELY NOT ADHERED TIGHTLY TO THE ADAPTOR. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM HAD BLOOD LEAKAGE BECAUSE THE CATHETER WAS LIKELY NOT ADHERED TIGHTLY TO THE ADAPTOR. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112216 BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 8233349

Patients

Seq Age Sex Outcome Treatment
1 Other