ENDOPATH ENDOSCOPIC LINEAR CUTTER
Report
- Report Number
- 1527736-1997-00493
- Event Type
- Malfunction
- Date Received
- April 3, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
H6; CODE 400: DAMAGED FIRING MECHANISM. FACILITY EXPERIENCED AN EVENT WITH YOUR ENDOPATH ENDOSCOPIC LINEAR CUTTER WHILE PERFORMING A UNK PROCEDURE. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #971710. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECH. ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, CARTRIDGE CONDITION, AND CARTRIDGE RETURN BATCH NUMBER; NA. INSTRUMENT NUMBER; 144. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, CONDITION OF PINION GEAR, AND CONDITION OF YOKE; GOOD. CONDITION OF SHORT RACK; BROKEN. ANALYSIS CONCLUSION: BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANISM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. SOME CONDITIONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYCLE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW RELOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE THE PRODUCTS.
THE DEVICE WAS USED DURING AN UNK PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE DEVICE WOULD NOT FIRE. THREE ADD'L RELOAD WERE TRIED WITH THE SAME OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC LINEAR CUTTER | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |