FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK SYRINGE LUER-LOK

MDR report key: 8323237 · Received February 8, 2019

Report

Report Number
1213809-2019-00161
Event Type
Malfunction
Date Received
February 8, 2019
Date of Event
January 18, 2019
Report Date
April 15, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096580
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: ONE PHOTO WAS RECEIVED DISPLAYING TWO LOOSE SYRINGES NEXT TO OPENED BLISTER PACKS OF BATCH #8210545 (P/N 309658). ONE OF THE PICTURES RECEIVED SHOWS THE SCALE OF THE SYRINGE SLIGHTLY OUT OF CENTER. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT THE ROOT CAUSE OF DEFECT FOUND -SCALE MISPLACED/ROLLED SCALE- IS RELATED TO THE PRINTING PROCESS. IT IS POSSIBLE THAT THIS DEFECT OCCURRED DURING THE SETUP OF THE PRINTER MACHINE AND IT IS POSSIBLE THAT LIMITED NUMBER OF SYRINGES WERE NOT PROPERLY REMOVED FROM THE PROCESS ¿ SCRAPPED. THE PERFORMANCE OF THE PRODUCT SHOULD NOT BE COMPROMISED BY ROLLED SCALE, THIS IMPERFECTION DOES NOT COMPROMISE THE PRODUCT¿S VOLUMETRIC ACCURACY.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD PLASTIPAK¿ SYRINGES LUER-LOK¿S HAD THE "ML-SCALE PRINTED ON THE LATERAL SIDE".

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD PLASTIPAK SYRINGES LUER-LOKS HAD THE "ML-SCALE PRINTED ON THE LATERAL SIDE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112379 BD PLASTIPAK SYRINGE LUER-LOK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS UNKNOWN 30382903096580

Patients

Seq Age Sex Outcome Treatment
1 Other