- Report Number
- 8010042-2019-00079
- Event Type
- Malfunction
- Date Received
- February 8, 2019
- Date of Event
- January 21, 2019
- Report Date
- April 15, 2019
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K151814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC (B)(4). CONTACT PERON: (B)(6). THE VENTILATOR WAS INVESTIGATED ON SITE BY OUR FIELD SERVICE ENGINEER (FSE). NO FAULT WAS FOUND WITH THE VENTILATOR, ALL TESTS PASSED AND NO PARTS WERE REPLACED. OUR FSE FOUND THAT A FILTER WAS USED ON THE INSPIRATORY SIDE. WHEN THE INSPIRATORY FILTER WAS REMOVED, THE ISSUE WITH THE 1-3 CMH2O PRESSURE SPIKES AT THE END OF INSPIRATION DISAPPEARED. THE FILTER WAS NOT MANUFACTURED OR DISTRIBUTED BY US. THE 1-3 CMH2O PRESSURE SPIKES WERE NOT RELATED TO LEAKAGE AS WAS REPORTED INITIALLY. EVALUATION OF THE RECEIVED DEVICE LOGS SHOW THAT PRE-USE CHECK PRIOR TO AND AFTER THE EVENT WAS SUCCESSFULLY. THERE IS NO ENTRY IN THE TECHNICAL LOG TO INDICATE A TECHNICAL FAILURE IN THE VENTILATOR. AFTER THE USER HAD REMOVED THE FILTER FROM THE INSPIRATORY SIDE NO FURTHER ISSUES HAVE BEEN REPORTED AND THE VENTILATOR HAS BEEN RETURNED BACK INTO CLINICAL USE.
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERON: (B)(4).
IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).
IT WAS REPORTED THAT WHILE THE VENTILATOR WAS CONNECTED TO A PATIENT, THERE WERE PRESSURE SPIKES LEADING TO PEAK PRESSURES RISES OF 2-3 CMH2O WHEN LEAKAGE WAS PRESENT. THE SPIKES DISAPPEARED ONCE LEAKAGE WAS CORRECTED. THERE WAS NO PATIENT HARM. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 112250 | SERVO-U | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 DA |