FDA Adverse Event Malfunction Summary report: N

HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT

MDR report key: 8319910 · Received February 8, 2019

Report

Report Number
3005180920-2019-00046
Event Type
Malfunction
Date Received
February 8, 2019
Date of Event
January 11, 2019
Report Date
March 6, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER: THE SPRING BALL PLUNGER IS CAME APART. IT IS POSSIBLE THIS OCCURRENCE WAS INFLUENCED BY VARIATION IN PRODUCTION/ASSEMBLY, HOWEVER THIS CANNOT BE CONFIRMED.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 08 FEBRUARY 2019. LOT 1851540: (B)(4). NO ANOMALIES FOUND RELATED TO THE ISSUE. TO DATE, NO SIMILAR EVENT ON THE SAME LOT HAS BEEN REGISTERED. OTHER LOT INVOLVED LOT 1210183C: (B)(4). NO ANOMALIES FOUND RELATED TO THE PROBLEM.

Description of Event or Problem · 1

THE BALL SPRING FELL OUT FROM 2 HANDLES. THE SURGERY WAS COMPLETED SUCCESSFULLY, WITH 5 MINUTES OF DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112220 HIP GENERAL OFFSET-30° BROACH HANDLE - RIGHT SURGICAL INSTRUMENT FOR HIP LXH MEDACTA INTERNATIONAL SA 1851540

Patients

Seq Age Sex Outcome Treatment
1 Other