FDA Adverse Event Malfunction Summary report: N

SHELL IMPACTION PLATFORM

MDR report key: 8319716 · Received February 8, 2019

Report

Report Number
3005985723-2019-00119
Event Type
Malfunction
Date Received
February 8, 2019
Date of Event
February 5, 2019
Report Date
May 7, 2019
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486022082
PMA / PMN Number
K121064
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

PRODUCT IDENTIFICATION: THE RETURNED DEVICE WAS CONFIRMED TO BE A TRIDENT INLINE OFFSET IMPACTOR, P/N 209830, L/N 32953, RM 267706. INSPECTION: VISUAL INSPECTION SHOWED THE DEVICE'S SHAFT BROKEN WHERE THE IMPACTION PLATFORM ASSEMBLES. PRODUCT HISTORY: REVIEW OF DEVICE HISTORY RECORDS SHOWS THE FOLLOWING NUMBER OF DEVICES ON THEIR GIVEN DATES WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES: 35 ON (B)(6) 2016, 24 OF 25 ON (B)(6) 2016, 37 ON (B)(6) 2016, 2 ON (B)(6) 2016, 2 ON (B)(6) 2016. THE ONE REJECTED DEVICE WAS DISPOSITIONED PER (B)(4) FOR AN OUT OF SPEC FEATURE UNRELATED TO THE FAILURE IN THIS INVESTIGATION. COMPLAINT HISTORY REVIEW: REVIEW OF COMPLAINTS WITHIN THE TRACKWISE DATABASE FOR P/N 209830, L/N 32953 SHOWS ZERO (0) COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD. CONCLUSION: FUNCTIONAL INSPECTION CONFIRMS ALLEGED FAILURE OF A BROKEN DEVICE.

Description of Event or Problem · 0

WHILE IMPACTING CUP ON THE FINAL STRIKE, THE END STRIKE PLATE FOR THE OFFSET IMPACTOR BROKE INTO PIECES. CASE TYPE: THA.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

WHILE IMPACTING CUP ON THE FINAL STRIKE, THE END STRIKE PLATE FOR THE OFFSET IMPACTOR BROKE INTO PIECES. CASE TYPE: THA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112445 SHELL IMPACTION PLATFORM STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 45011215 00848486022082

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization