FDA Adverse Event Malfunction Summary report: N

NIT-OCCLUD PDA

MDR report key: 8316341 · Received February 7, 2019

Report

Report Number
3005704822-2018-00001
Event Type
Malfunction
Date Received
February 7, 2019
Date of Event
January 1, 2018
Report Date
February 7, 2019
Manufacturer
PFM MEDICAL MEPRO GMBH
Product Code
MAE
UDI-DI
04042301160034
PMA / PMN Number
P120009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS PROVIDED AND WAS USED TO CONDUCT AN INVESTIGATION. THE DEVICE HISTORY RECORD (DHR) HAS BEEN REVIEWED AND SHOWED NO DEVIATIONS. THE COIL, ITEM 145116, WAS STILL MOUNTED ON THE CARRIER AND WITH THE PROXIMAL PART STILL IN THE TRANSPORTATION SHEATH. THE CATHETER WAS CONTAMINATED WITH BLOOD AND NO LONGER USABLE. IF THE CATHETER HAS NOT BEEN FLUSHED ENOUGH DURING IMPLANTATION, A GRAFT MAY FORM WHICH THEN MAKES THE REASON FOR THE COMPLAINT PLAUSIBLE. TO RECONSTRUCT THE REASON FOR THE COMPLAINT, A NEW IMPLANTATION CATHETER WAS USED. THE COIL COULD BE EASILY PUSHED THROUGH THE CATHETER, CONFIGURED, AND WITHDRAWN WITHOUT PROBLEMS. NO FURTHER SIMILAR COMPLAINTS WERE REPORTED TO MEPRO IN THE PAST. BASED ON PREVIOUS EXPERIENCE, INSUFFICIENT RINSING CANNOT BE RULED OUT. THE COMPLAINT IS NOT JUSTIFIED.

Description of Event or Problem · 1

THE COIL WENT OUT OF THE TRANSPORTATION SHEATH. DOCTOR'S REPORT: "AFTER INTRODUCING THE COIL, WE TRIED TO REPOSITION IT. AT THE FIRST TRIAL OF REPOSITION THE LOOPS STUCKED ON THE CATHETER AND REFUSED TO GO BACK INSIDE THE CATHETER OR REFIGURED AGAIN AT ALL. AFTER FLUSHING THE SYSTEM SEVERAL TIMES, WE TRIED TO PULL IT HARDER TO LET IT MOVE, THEN IT'S WENT OUT OF THE TRANSPORTATION SHEATH." COULD RESULT IN EMBOLISM OF THE DEVICE IS OCCURRED AGAIN REQUIRED INTERVENTION TO REMOVE THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
112120 NIT-OCCLUD PDA PDA OCCLUDER MAE PFM MEDICAL MEPRO GMBH 145116 1021317 04042301160034

Patients

Seq Age Sex Outcome Treatment
1 Other