FDA Adverse Event Summary report: N

QUICK CHECK ONE

MDR report key: 83137 · Received April 10, 1997

Report

Report Number
2027242-1997-90124
Date Received
April 10, 1997
Date of Event
March 12, 1997
Report Date
March 12, 1997
Manufacturer
DIAGNOSTIC SOLUTIONS, INC.
Product Code
CFR
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED OUT OF RANGE NORMAL CONTROL SOLUTION TEST RESULTS WITH TEST STRIPS. ON 3/12/97, THE CUSTOMER OPENED THE SECOND BOTTLE FROM A BOX OF FIFTY AND PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 170 MG/DL VERSUS A NORMAL CONTROL SOLUTION RANGE OF 110-166 MG/DL. THE CUSTOMER CALLED THE CO CUSTOMER SUPPORT LINE AND, WITH THE REP, PERFORMED A NORMAL CONTROL SOLUTION TEST. THE RESULT WAS 174 MG/DL. THE CONTROL SOLUTION, LOT# V1236, EXPIRATION 9/98, WAS OPENED TODAY AND WAS SHAKEN VIGOROUSLY BEFORE THE SAMPLE WAS APPLIED. THE CUSTOMER WAS UNABLE TO PERFORM A CHECK STRIP TEST WITH THE REP BECAUSE HE HAD MISPLACED THE TEST STRIP. THE CUSTOMER STATED THAT THE FIRST BOTTLE PERFORMED ACCURATELY. THE DESICCANT IS IN THE OPEN BOTTLE. THE CUSTOMER STORES THE TEST STRIPS IN THE BEDROOM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUICK CHECK ONE BLOOD GLUCOSE TEST STRIPS FOR USE WITH ONE TOUCH METERS CFR DIAGNOSTIC SOLUTIONS, INC. * 1L513B

Patients

Seq Age Sex Outcome Treatment
1 *