FDA Adverse Event Malfunction Summary report: N

ENDO GIA ULTRA

MDR report key: 8312886 · Received February 6, 2019

Report

Report Number
2647580-2019-00727
Event Type
Malfunction
Date Received
February 6, 2019
Date of Event
December 17, 2018
Report Date
March 7, 2019
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDW
UDI-DI
10884523003543
PMA / PMN Number
K111825
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. INITIAL VISUAL INSPECTION OF THE INSTRUMENT NOTED NO ABNORMALITIES. FUN CTIONALLY, THE INSTRUMENT WAS LOADED WITH AN ARTICULATING VASCULAR/MEDIUM RELOAD. IT WAS OBSERVED THAT THE INSTRUMENT WAS ABLE TO BE CYCLED WITHOUT PRESSING THE GREEN FIRING BUTTON. THE INSTRUMENT WAS DISASSEMBLED FOR VISUALIZATION OF INTERNAL COMPONENTS. THIS EXAMINATION FOUND THAT THE GRASPING MECHANISM WAS FRACTURED. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. REPLICATION OF THE GRASPING MECHANISM DAMAGE MAY OCCUR WHEN AN ATTEMPT IS MADE TO PRY OR PUSH THE TRIGGER HANDLE OPEN WHILE THE GRASPER IS ENGAGED. THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS MISUSE OF THE PRODUCT WHICH CAUSED OR CONTRIBUTED TO THE REPORTED CONDITION. NO FURTHER ACTIONS HAVE BEEN DEEMED NECESSARY AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPY, UPON GRASPING THE TISSUE, WHEN THEY INSERTED THE INITIALLY LOADED HANDLE WITH RELOAD, THE DEVICE WOULD NOT FIRE, THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON AND THE HANDLE DID NOT SQUEEZE. THE HANDLE WOULD NOT RELEASE THE RELOAD. THEY REPLACED BOTH THE HANDLE AND RELOAD TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105264 ENDO GIA ULTRA STAPLE, IMPLANTABLE GDW US SURGICAL PUERTO RICO EGIAUSTND P8J1452X 10884523003543

Patients

Seq Age Sex Outcome Treatment
1