BD DISCARDIT II SYRINGE W/O NEEDLE
Report
- Report Number
- 3002682307-2019-00128
- Event Type
- Malfunction
- Date Received
- February 6, 2019
- Date of Event
- January 15, 2019
- Report Date
- March 18, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: DHR: ¿1806152: BD HAS REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QN NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2011 (JUNE 7TH - 8TH, 2018). SYRINGES WERE ASSEMBLED IN MACHINE IN LOT #8155592, #8148567. RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARREL LOTS #8156556, #8149705, #8142763, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #8156560, #8142767, #8144709, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. 1807214: BD HAS REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QN NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2022 (JUNE 21ST - 22ND, 2018). SYRINGES WERE ASSEMBLED IN MACHINE IN LOT #8197930. RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. BD HAS ALSO REVIEWED THE BARREL LOTS #8198680, #8185517, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #8198684, #8185526, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. 1807134: BD HAS REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QN NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2022 (JUNE 11TH - 12TH, 2018). SYRINGES WERE ASSEMBLED IN MACHINE IN LOT #8186940. RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARREL LOTS #8185517, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #8163671, #8185526, AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. BD HAS BEEN PROVIDED WITH THE AFFECTED SAMPLES. AFTER THE EVALUATION OF THE RETURNED SAMPLES, BD CONFIRMED THE REPORTED ISSUE, AND CONCLUDED THAT THE WHITE PARTICLES INSIDE THE SYRINGE WERE COMPOSED BY LUBRICANT FROM THE SYRINGE BARREL. CONCLUSION(S): PARTICLES COMPOSED BY LUBRICANT, WHICH IS INCLUDED IN THE PLASTIC FORMULATION AND IT IS INHERENT TO THE DESIGN AND FUNCTION OF 2 PIECE SYRINGES. BD HAS INITIATED THE APPROPRIATE CORRECTIVE AND PREVENTIVE ACTIONS FOR THIS ISSUE. CAPA #207816.
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD DISCARDIT¿ II SYRINGES W/O NEEDLES WERE NOTED TO CONTAIN FOREIGN MATTER CONTAMINATION.
(B)(6). THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1806152, MEDICAL DEVICE EXPIRATION DATE: 05/31/2023, DEVICE MANUFACTURE DATE: 05/31/2018. MEDICAL DEVICE LOT #: 1807214, MEDICAL DEVICE EXPIRATION DATE: 06/30/2023, DEVICE MANUFACTURE DATE: 07/17/2018. MEDICAL DEVICE LOT #: 1807134, MEDICAL DEVICE EXPIRATION DATE: 06/30/2023, DEVICE MANUFACTURE DATE: 06/20/2018. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD DISCARDIT¿ II SYRINGES W/O NEEDLES WERE NOTED TO CONTAIN FOREIGN MATTER CONTAMINATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106680 | BD DISCARDIT II SYRINGE W/O NEEDLE | SYRINGE | FMF | BECTON DICKINSON, S.A. | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |