FDA Adverse Event Injury Summary report: N

MINIMAX CEMENTLESS ANATOMICAL STEM RIGHT SIZE 6

MDR report key: 8310185 · Received February 6, 2019

Report

Report Number
3005180920-2019-00024
Event Type
Injury
Date Received
February 6, 2019
Date of Event
January 7, 2019
Report Date
February 6, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030803260
PMA / PMN Number
K170845
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 05 FEBRUARY 2019: LOT 181832: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 11-JUL-2018. EXPIRATION DATE: 2023-07-02. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR ON JANUARY 25, 2019. FEMORAL HEAD REVISION FEW WEEKS AFTER TOTAL HIP ARTHROPLASTY. ACCORDING TO THE REPORT, THE STEM SUBSIDED BUT THIS CANNOT BE ASSESSED ON THE BASIS OF A SINGLE X-RAY. EARLY SUBSIDENCE MAY BE ASSOCIATED WITH INSUFFICIENT PRIMARY STABILITY ACHIEVED AT THE TIME OF OPERATION. NO CONCLUSION CAN BE DRAWN ON THE BASIS OF AVAILABLE INFORMATION, BUT THERE IS NO REASON TO SUSPECT A FAULTY DEVICE.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF PAIN CAUSED BY A SUBSIDED STEM AFTER ABOUT 18 DAYS FROM PRIMARY. THE SURGEON CHOSE TO LEAVE THE STEM IN PLACE BUT REVISED HEAD AND ADDED A SLEEVE OPTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
102716 MINIMAX CEMENTLESS ANATOMICAL STEM RIGHT SIZE 6 CEMENTLESS ANATOMICAL STEM LZO MEDACTA INTERNATIONAL SA 181832 07630030803260

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R