FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 8308617 · Received February 5, 2019

Report

Report Number
1645337-2019-08339
Event Type
Injury
Date Received
February 5, 2019
Date of Event
January 4, 2019
Report Date
January 9, 2019
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001232
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 3/6/2019, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. ON 3/26/2019, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.1 CM WITHIN A CREASE ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. A MANUFACTURING RECORD EVALUATION (MRE) OF LOT NUMBER 6840978 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 5/24/2019, MENTOR BECAME AWARE THAT THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH THE FOLLOWING DEVICES: (RIGHT) 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE, CATALOG: 3501640, LOT: 7646565, SN: (B)(4); AND (LEFT) 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE, CATALOG: 3501640, LOT: 7598884, SN: (B)(4). MENTOR ALSO BECAME AWARE THAT THE PATIENT ALSO EXPERIENCED CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN ON THE LEFT SIDE (REPORTED UNDER 1645337-2019-14181). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT MEDICAL PRODUCTS: 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE, CATALOG: 3501640, LOT: 6840978, SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT WHO UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 275CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESES, EXPERIENCED RIGHT SIDE DEFLATION POST-OPERATIVELY. AS A RESULT, THE PATIENT WILL UNDERGO REMOVAL ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96931 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 6840978 00081317001232

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention