FDA Adverse Event Malfunction Summary report: N

METRIX AIR

MDR report key: 8306685 · Received February 5, 2019

Report

Report Number
1000113657-2019-00097
Event Type
Malfunction
Date Received
February 5, 2019
Date of Event
January 11, 2019
Report Date
March 6, 2019
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007720
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) INTERNAL REPORT #(B)(4) RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58-USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP UDI# (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER (SEVERAL ATTEMPTS) IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH THE CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CUSTOMER TO CONTACT CUSTOMER CARE.

Description of Event or Problem · 0

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS OF 223, 284, 303, 157 AND 177 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 140 MG/DL. CUSTOMER IS USING THE CORRECT TESTING TECHNIQUE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON 01/11/2019, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/29/2020 AND OPEN VIAL DATE IS 01/04/2019. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1:223MG/DL DATE:(B)(6) 2019, TIME:10:47PM FASTING, RESULT 2:284MG/DL DATE:(B)(6) 2019, TIME:11:53AM FASTING, RESULT 3:303MG/DL DATE:(B)(6) 2019, TIME:1:37AM FASTING, RESULT 4:157MG/DL DATE:(B)(6) 2019, TIME:9:55PM FASTING, RESULT 5:177MG/DL DATE:(B)(6) 2019, TIME:12:07PM FASTING.

Additional Manufacturer Narrative · 1

INTERNAL REPORT # (B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-58-USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. TEST STRIP UDI# (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER (SEVERAL ATTEMPTS) IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH THE CUSTOMER AT THIS TIME. PRODUCT NOTIFICATION LETTER SENT TO CUSTOMER TO CONTACT CUSTOMER CARE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS OF 223, 284, 303, 157 AND 177 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 100 - 140 MG/DL. CUSTOMER IS USING THE CORRECT TESTING TECHNIQUE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(4) 2019, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/29/2020 AND OPEN VIAL DATE IS (B)(6) 2019. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1:223MG/DL DATE: (B)(6) 2019 TIME:10:47PM FASTING; RESULT 2:284MG/DL DATE: (B)(6) 2019 TIME:11:53AM FASTING; RESULT 3:303MG/DL DATE: (B)(6) 2019 TIME:1:37AM FASTING; RESULT 4:157MG/DL DATE: (B)(6) 2019 TIME:9:55PM FASTING; RESULT 5:177MG/DL DATE: (B)(6) 2019 TIME:12:07PM FASTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
99323 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MV2909 00021292007720

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY| SECOND THERAPY