HANDICARE STAIRLIFTS 2000
Report
- Report Number
- 3011268530-2019-00005
- Event Type
- Malfunction
- Date Received
- February 5, 2019
- Date of Event
- July 1, 2018
- Report Date
- January 20, 2020
- Manufacturer
- HANDICARE ACCESSIBILITY UK LTD
- Product Code
- PCD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
(B)(4). CORRECTION: JULY 2018, A NEW UNIVERSAL END STOP WAS LAUNCHED. FURTHER IMPROVEMENTS ARE BEING MADE UNDER THE SAME ECN TO IMPROVE FITABILITY. ALL RECORDED FAILURES OF THE END STOPS ARE WITH THE OLD DESIGNS MADE PRIOR TO JULY 2018. 25 PRODUCTION SAMPLES OF THE UNIVERSAL END STOP WERE TESTED AND THEY SUCCESSFULLY STOPPED THE POWER PACK LOADED WITH 342.5KG WITH ALL ELECTRICAL SAFETIES REMOVED. THE WELDING WAS INSPECTED BY R&D AND QUALITY. FURTHER TESTING WAS PERFORMED ON THE UNIVERSAL END STOP OVER 66,530 CYCLES AND PASSED WITH NO REPORTED ISSUES. ALL REPORTED FAILURES HAVE OCCURED WITH AN OLDER DESIGN. MANUFACTURER INTRODUCED THE UNIVERSAL END STO IN JULY 2018 AND FAILURES ARE REPORTED.
MECHANICAL END STOP WELD BROKE OF THE RAIL AND THE PACK CAME OFF THE RAIL. NO INJURIES REPORTED.
RETROSPECTIVE MDR REPORTING.
MECHANICAL END STOP WELD BROKE OF THE RAIL AND THE PACK CAME OFF THE RAIL. NO INJURIES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100456 | HANDICARE STAIRLIFTS 2000 | STAIRLIFTS | PCD | HANDICARE ACCESSIBILITY UK LTD | INTERNAL CURVE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |