PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2019-00795
- Event Type
- Injury
- Date Received
- February 4, 2019
- Date of Event
- January 10, 2019
- Report Date
- March 27, 2019
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
INTERNAL FILE NUMBER - (B)(4). CORRECTION: LOT NUMBER UPDATED FROM 8071643 TO 8092141. EVALUATION SUMMARY: VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. IT SHOULD BE NOTED THAT THE DEPLOYMENT SEQUENCE IN THE PERCLOSE PROGLIDE INSTRUCTIONS FOR USE, STATES: PERFORM A FEMORAL ANGIOGRAM TO VERIFY THE LOCATION OF THE PUNCTURE SITE. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. (B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL TWO PROGLIDE DEVICES REFERENCED "IS" BEING FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.
IT WAS REPORTED THAT SUTURE PLACEMENT WITH THREE PROGLIDE DEVICES WERE ATTEMPTED IN THE RIGHT COMMON FEMORAL ARTERY VIA 6FR SHEATH USING THE PRECLOSE TECHNIQUE PRIOR TO AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, WHEN THE PLUNGER WAS RETRIEVED, NO SUTURES WERE NOTICED; A CUFF MISS OCCURRED WITH THE THREE PROGLIDE DEVICES. TWO ADDITIONAL PROGLIDE DEVICES WERE USED FOR SUCCESSFUL SUTURE PLACEMENT USING THE PRECLOSE TECHNIQUE. THE SHEATH WAS UPSIZED TO 16FR AND THE AAA PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED USING THE PRE-PLACED SUTURES FROM THE PROGLIDE DEVICES. A FEMORAL ANGIOGRAM WAS NOT TAKEN. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96102 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 8092141 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | SHEATH: 6FRHEPARIN| SHEATH: 6FRHEPARIN |