FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 8301034 · Received February 4, 2019

Report

Report Number
2024168-2019-00795
Event Type
Injury
Date Received
February 4, 2019
Date of Event
January 10, 2019
Report Date
March 27, 2019
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

INTERNAL FILE NUMBER - (B)(4). CORRECTION: LOT NUMBER UPDATED FROM 8071643 TO 8092141. EVALUATION SUMMARY: VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. IT SHOULD BE NOTED THAT THE DEPLOYMENT SEQUENCE IN THE PERCLOSE PROGLIDE INSTRUCTIONS FOR USE, STATES: PERFORM A FEMORAL ANGIOGRAM TO VERIFY THE LOCATION OF THE PUNCTURE SITE. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. (B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE ADDITIONAL TWO PROGLIDE DEVICES REFERENCED "IS" BEING FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT SUTURE PLACEMENT WITH THREE PROGLIDE DEVICES WERE ATTEMPTED IN THE RIGHT COMMON FEMORAL ARTERY VIA 6FR SHEATH USING THE PRECLOSE TECHNIQUE PRIOR TO AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, WHEN THE PLUNGER WAS RETRIEVED, NO SUTURES WERE NOTICED; A CUFF MISS OCCURRED WITH THE THREE PROGLIDE DEVICES. TWO ADDITIONAL PROGLIDE DEVICES WERE USED FOR SUCCESSFUL SUTURE PLACEMENT USING THE PRECLOSE TECHNIQUE. THE SHEATH WAS UPSIZED TO 16FR AND THE AAA PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED USING THE PRE-PLACED SUTURES FROM THE PROGLIDE DEVICES. A FEMORAL ANGIOGRAM WAS NOT TAKEN. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96102 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 8092141

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6FRHEPARIN| SHEATH: 6FRHEPARIN