SERVO-I
Report
- Report Number
- 8010042-2019-00059
- Event Type
- Malfunction
- Date Received
- February 1, 2019
- Date of Event
- January 14, 2019
- Report Date
- August 16, 2019
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K123149
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- RISK MANAGER
Narratives
THE VENTILATOR WAS INVESTIGATED ON-SITE BY OUR FIELD SERVICE ENGINEER. THE REPORTED VENTILATOR SHUTDOWN COULD NOT BE DUPLICATED AND THE VENTILATOR PASSED PRE-USE CHECKS. NO PARTS WERE REPLACED. VENTILATOR LOGS WERE DOWNLOADED. LOGS CONTAINS A SHUTDOWN NOT PRECEDED BY A CORRECT STANDBY. THERE WAS ONLY ONE LOG POST; STANDBY BUTTON ACTIVATED. (A NORMAL SHUTDOWN IN THE LOG IS PRECEDED BY ENTRIES: STANDBY BUTTON ACTIVATED, STANDBY DIALOG CONFIRMED AND STANDBY). THE VENTILATOR PASSED PRE-USE CHECK PRIOR THE REPORTED EVENT AND THE TECHNICAL LOG DOES NOT CONTAIN ANY TECHNICAL ERROR CODES CONNECTED TO THE SHUTDOWNS. A SHUTDOWN WITHOUT ANY ENTRY IN THE TECHNICAL LOG IS CONSIDERED TO BE A COMPLETE POWERING OFF OF THE VENTILATOR USING THE ON/OFF SWITCH. THIS IS CONSIDERED AS A NORMAL SHUTDOWN AND NO OTHER ALARMS ARE GENERATED. THE VENTILATOR CANNOT SHUTDOWN BY ITSELF WHEN THE ON/OFF SWITCH IS IN THE ON POSITION. THE CONCLUSION IS THAT THE SHUTDOWN OCCURRED WHILE THE ON/OFF SWITCH WAS IN THE OFF POSITION. THE CAUSE COULD HAVE BEEN TURNING OFF OF THE VENTILATOR INADVERTENTLY OR A DELAY OF THE VENTILATOR TO ATTAIN THE OFF STATE. TWO PREVENTIVE MEASURES CONSISTING OF A NEW TOGGLE SWITCH TO PREVENT INADVERTENT SWITCH OFF AND SOFTWARE MONITORING OF THE ON/OFF CIRCUITRY TO DETECT AN INACCURATE ON-OFF STATE LIKE THE DELAY IN SWITCHING OFF WERE IMPLEMENTED IN (B)(6). THE VENTILATOR IN THIS REPORT WAS MANUFACTURED BEFORE THESE MEASURES.
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERON: (B)(4).
IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: 1904MCC0282.
IT WAS REPORTED THAT THE VENTILATOR SHUTDOWN DURING PATIENT TREATMENT. THERE WAS NO PATIENT HARM. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 88828 | SERVO-I | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |