LINK ENDO-MODEL® ROTATION KNEE JOINT PROSTHESIS - CEMENTABLE
Report
- Report Number
- 3004371426-2019-00069
- Event Type
- Injury
- Date Received
- February 1, 2019
- Date of Event
- July 11, 2016
- Report Date
- June 23, 2017
- Manufacturer
- WALDEMAR LINK GMBH & CO. KG
- Product Code
- KRO
- PMA / PMN Number
- K143179
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
ACCORDING TO THE USER REPORT, THE ARTICLE IN QUESTION IS (B)(4),WHICH WAS ALSO STATED IN THE COMPLAINT REPORT, BUT THE TINBN KNEE WAS RETURNED WITH ARTICLE NUMBERS 15-9024/12 AND SN (B)(4) THE ARTICLE (B)(4) IS FILED AND INVESTIGATED UNDER (B)(4) (MDR 3004371426-2019-00069). THE ARTICLE (B)(4) AND SN (B)(4) IS NOW FILED AND INVESTIGATED UNDER (B)(4) (MDR 3004371426-2019-00127).
TRANSLATION: ON (B)(6) 2016 THE PATIENT WAS EXAMINED DUE TO INCREASED PAIN AFTER THE LAST OPERATION WITH KNEE PROSTHESIS CHANGE ON (B)(6) 2010. THE KNEE JOINT BLOCKS WHEN STANDING UP AND STARTING UP. THE KNEE IS OVERSTRETCHED. THERE WAS A MEDIOLATERAL FOLDABILITY WITH UNPHYSIOLOGICAL JOINT PLAY IN THE SOCKET MECHANISM. THE RIGHT KNEE JOINT CAN ONLY BE BENT AGAINST SLIGHT RESISTANCE, A MECHANISM THAT RUNS WITH DIFFICULTY. AFTER EXPLANTATION OF THE PROSTHESIS ON (B)(6) 2016, A FRACTURE IN THE COUPLING MECHANISM/BUSHING DAMAGE TO THE PROSTHESIS WAS OBSERVED."
ACCORDING TO THE USER REPORT, THE ARTICLE IN QUESTION IS 15-8030/11 SN (B)(4), WHICH WAS ALSO STATED IN THE COMPLAINT REPORT, BUT THE TINBN KNEE WAS RETURNED WITH ARTICLE NUMBERS 15-9024/12 AND SN (B)(4). THE REPORT IS DELAYED DUE TO INCONSISTENCIES WITH REGARD TO FOREIGN EVENT REPORTING TO FDA. AFTER RE-EVALUATION OF THE COMPLAINT WE DETERMINED THAT IT IS REPORTABLE. WE HAVE ADDRESSED THE INCONSISTENCY IN REPORTING FOREIGN EVENTS TO FDA THROUGH CAPA-(B)(4).
TRANSLATION: ON (B)(6) 2016 THE PATIENT WAS EXAMINED DUE TO INCREASED PAIN AFTER THE LAST OPERATION WITH KNEE PROSTHESIS CHANGE ON (B)(6) 2010. THE KNEE JOINT BLOCKS WHEN STANDING UP AND STARTING UP. THE KNEE IS OVER STRETCHABLE. THERE WAS A MEDIOLATERAL FOLDABILITY WITH UNPHYSIOLOGICAL JOINT PLAY IN THE SOCKET MECHANISM. THE RIGHT KNEE JOINT CAN ONLY BE BENT AGAINST SLIGHT RESISTANCE, A MECHANISM THAT RUNS WITH DIFFICULTY. AFTER EXPLANTATION OF THE PROSTHESIS ON (B)(6) 2016, A FRACTURE IN THE COUPLING MECHANISM/BUSHING DAMAGE TO THE PROSTHESIS WAS OBSERVED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 89831 | LINK ENDO-MODEL® ROTATION KNEE JOINT PROSTHESIS - CEMENTABLE | ROTATING HINGED TOTAL KNEE PROSTHESIS | KRO | WALDEMAR LINK GMBH & CO. KG | 15-8030/11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |