FDA Adverse Event Malfunction Summary report: N

11.5MM MEDULLARY REAMER HEAD

MDR report key: 8294785 · Received January 31, 2019

Report

Report Number
2939274-2019-56190
Event Type
Malfunction
Date Received
January 31, 2019
Report Date
January 15, 2019
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTO
UDI-DI
10886982194237
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10: ADDITIONAL NARRATIVE: PART 352.115, LOT 2082345: MANUFACTURING SITE: SELZACH. SUPPLIER: (B)(4). RELEASE TO WAREHOUSE DATE: DECEMBER 22, 2003. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. H3, H6: A PRODUCT INVESTIGATION WAS COMPLETED: THE SYNREAM REAMER HEAD WAS RECEIVED WITH THE REAMER HEAD SMOOTH TO THE TOUCH. THE REAMER HEAD CUTTING SURFACES ARE DULL COMPARED TO A NEW REAMER HEAD WHICH WOULD HAVE TO BE HANDLED WITH A LOT MORE CARE. THIS IS CONSISTENT WITH THE REPORTED COMPLAINT CONDITION, THUS CONFIRMING THE COMPLAINT. THE COMPLAINT CONDITION IS CONFIRMED AS THE SYNREAM REAMER HEAD WAS RECEIVED WITH THE REAMER HEAD DULL. AFTER A VISUAL INSPECTION, IT IS DETERMINED THAT THE REAMER HEAD IS WORN FROM REPEATED USE AND SERVICING; THEREFORE, FURTHER INVESTIGATION FOR THE REPORTED COMPLAINT DEVICE IS NOT REQUIRED. DURING THE INVESTIGATION, NO PRODUCT DESIGN OR MANUFACTURING ISSUES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION; THEREFORE, FURTHER CORRECTIVE AND/OR PREVENTIVE ACTION IS NOT REQUIRED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2019, ONE (1) FLEXIBLE REAMERS FOR INTRAMEDULLARY NAILS INSTRUMENT SET GRAPHIC CASE PLASTIC DIVIDERS WERE PEELING OFF AND EIGHT (8) MEDULLARY REAMER HEADS WERE DULL. PROCEDURE AND PATIENT INVOLVEMENT ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 11.5MM MEDULLARY REAMER HEAD. THIS IS REPORT 7 OF 8 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86570 11.5MM MEDULLARY REAMER HEAD REAMER HTO WRIGHTS LANE SYNTHES USA PRODUCTS LLC 352.115 2082345 10886982194237

Patients

Seq Age Sex Outcome Treatment
1