MDR report key: 829338 · Received March 16, 2007

Report

Report Number
MW1042254
Event Type
Injury
Date Received
March 16, 2007
Date of Event
March 2, 2007
Report Date
March 9, 2007
Manufacturer
MOHAWK DENTAL SUPPLY CO., INC.
Product Code
EGY
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

ACCORDING TO THE FIRE DEPT. INVESTIGATORS, AN ELECTRICAL MALFUNCTION IN A DENTAL X-RAY PROCESSING MACHINE CAUSED THE WIRING WITHIN THE MACHINE TO OVERHEAT. THIS OVERHEATED WIRING IGNITED PART OF THE MACHINE AND NEARBY COMBUSTIBLES. THE X-RAY PROCESSING MACHINE WAS A DENT X 810 PROCESSOR MACHINE, INSTALLED BY MOHAWK DENTAL SUPPLY CO. ON 10/24/05. THE ROOM THE EQUIPMENT WAS LOCATED IN WAS CONSUMED BY EXTENSIVE FIRE DAMAGE. SMOKE FROM THE FIRE EXTENDED TO OTHER AREAS IN THE BUILDING AND WAS DIFFICULT TO VENTILATE, NEARLY REQUIRED EVACUATION OF THE ATTACHED NURSING FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DENT X FILM PROCESSOR EGY MOHAWK DENTAL SUPPLY CO., INC. 810 PROCESSOR MACH *

Patients

Seq Age Sex Outcome Treatment
1 NA Life Threatening| O