FDA Adverse Event Malfunction Summary report: N

HANIDCARE STAIRLIFTS

MDR report key: 8292942 · Received January 31, 2019

Report

Report Number
3011268530-2019-00004
Event Type
Malfunction
Date Received
January 31, 2019
Date of Event
December 1, 2017
Report Date
January 20, 2020
Manufacturer
HANDICARE ACCESSIBILITY UK LTD
Product Code
PCD
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

APPROXIMATELY 5429 STAIRLIFTS WERE SOLD BY MANUFACTURER FROM MAY 2017 TO JANUARY 2019. THIS WAS THE ONLY CUSTOMER COMPLAINT REGARDING THE TOGGLE MECHANISM. ROOT CAUSE: INITIAL DESIGN TESTING FOUND THAT THE TOGGLE OPERATED SATISFACTORILY FOR THE FIRST 17,000 CYCLES, BUT THEN THE BLUE TOGGLE MOULDING WOULD NOT RE-CENTRE ITSELF AFTER ACTIVATION. A BROKEN OR FATIGUED SPRING WAS SUSPECTED BUT ON DISMANTLING THE ARM IT BECAME CLEAR THAT THE TOGGLE MOULDING WAS BINDING ON THE ARM CASTING. THE SPRING WAS NOT DAMAGED. A VERY SMALL AMOUNT OF GREASE WAS APPLIED TO THE OUTER DIAMETER OF THE BLUE MOULDING WHERE IT RUBS ON THE INNER PAINTED BORE OF THE TOGGLE HOUSING. THIS CURED THE PROBLEM AND THE TOGGLE CONTINUED TO 100,000 CYCLES WITHOUT FURTHER INCIDENT. CORRECTIVE ACTION: A REVIEW OF THE SERVICE AND MAINTENANCE PROCEDURE FOUND THAT THERE IS NO INSTRUCTION TO LUBRICATE THE TOGGLE MECHANISM DURING REGULAR SERVICING. MANUFACTURER WILL CONTINUE TO MONITOR THESE TYPE OF EVENTS.

Description of Event or Problem · 0

THE SEAT TOGGLE STUCK THE USER AND SHE WAS TRYING TO GET ON THE LIFT AND IT MOVED. SHE FELL, CUT HER ARM AND BROUGHT HER DOWN THE STAIRS ON HER BACK. USER SLID OFF THE SEAT DOWN THE SIDE OF IT AND THE LIFT CONTINUED TO MOVE GOING DOWN. USER WEDGED DOWN THE SIDE AND IT STARTED TO DRAG HER DOWN ON HER BACK, USER SCREAMED AND A LADY IN THE HOUSE TRIED TO HELP,WHILE THE LIFT WAS ABOUT HALF WAY DOWN. THE LIFT WAS STOPPED BY THE CUSTOMER THROWING HER ARM UP TO KNOCK THE STAIRLIFT ARMS UP AS SHE'S AWARE THAT CUTS THE LIFT OUT,. THIS IS WHAT WE BELIEVE THINK CAUSED THE CUT ON HER ARM. THE LIFT WAS MOVING BY ITSELF UP TO A WEEK BEFORE THE INCIDENT, BUT IN THE LAST COUPLE WEEKS CUSTOMER TOLD SON THAT IF YOU PUSH THE LIFT TO GO DOWN AND LET GO IT CARRIED ON GOING, AND THE SON SAW THIS FOR HIMSELF AT A LATER DATE. AN ENGINEER HAD GONE OUT TO CHANGE THE BATTERY AND THE TIGHTENED THE TOGGLE, IT DIDN'T SPRING BACK 100 % INSTANTLY AS IT SHOULD, AND IT WAS STILL A BIT SLACK. THE ENGINEER ALSO MENTIONED THAT THE SAFETY DEVICE WHEREBY IF YOU LIFT THE ARM UP IT CUTS THE LIFT HADN'T BEEN CONNECTED IT HAD BEEN CONNECTED BEFORE THIS INCIDENT. THE USER WAS BRUISED ALL OVER BUT WAS NOT HOSPITALIZED.

Additional Manufacturer Narrative · 1

FILING RETROSPECTIVE MEDICAL DEVICE REPORTS.

Description of Event or Problem · 1

THE SEAT TOGGLE STUCK THE USER AND SHE WAS TRYING TO GET ON THE LIFT AND IT MOVED. SHE FELL, CUT HER ARM AND BROUGHT HER DOWN THE STAIRS ON HER BACK. USER SLID OFF THE SEAT DOWN THE SIDE OF IT AND THE LIFT CONTINUED TO MOVE GOING DOWN. USER WEDGED DOWN THE SIDE AND IT STARTED TO DRAG HER DOWN ON HER BACK, USER SCREAMED AND A LADY IN THE HOUSE TRIED TO HELP, WHILE THE LIFT WAS ABOUT HALF WAY DOWN. THE LIFT WAS STOPPED BY THE CUSTOMER THROWING HER ARM UP TO KNOCK THE STAIRLIFT ARMS UP AS SHE'S AWARE THAT CUTS THE LIFT OUT. THIS IS WHAT WE BELIEVE THINK CAUSED THE CUT ON HER ARM. THE LIFT WAS MOVING BY ITSELF UP TO A WEEK BEFORE THE INCIDENT, BUT IN THE LAST COUPLE WEEKS CUSTOMER TOLD SON THAT IF YOU PUSH THE LIFT TO GO DOWN AND LET GO IT CARRIED ON GOING, AND THE SON SAW THIS FOR HIMSELF AT A LATER DATE. AN ENGINEER HAD GONE OUT TO CHANGE THE BATTERY AND THE TIGHTENED THE TOGGLE, IT DIDN'T SPRING BACK 100 % INSTANTLY AS IT SHOULD, AND IT WAS STILL A BIT SLACK. THE ENGINEER ALSO MENTIONED THAT THE SAFETY DEVICE WHEREBY IF YOU LIFT THE ARM UP IT CUTS THE LIFT HADN'T BEEN CONNECTED IT HAD BEEN CONNECTED BEFORE THIS INCIDENT. THE USER WAS BRUISED ALL OVER BUT WAS NOT HOSPITALIZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
86700 HANIDCARE STAIRLIFTS STAIRLIFTS PCD HANDICARE ACCESSIBILITY UK LTD 950PLUS

Patients

Seq Age Sex Outcome Treatment
1