FDA Adverse Event Injury Summary report: N

CONSTRAINED LINER WITH CONSTRAINING RING

MDR report key: 8292097 · Received January 31, 2019

Report

Report Number
0001822565-2019-00353
Event Type
Injury
Date Received
January 31, 2019
Date of Event
January 15, 2019
Report Date
August 12, 2019
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWZ
PMA / PMN Number
K071718
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CONCOMITANT MEDICAL PRODUCTS: UNKNOWN STEM - UNKNOWN PART AND LOT, UNKNOWN CUP - UNKNOWN PART AND LOT. REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS SHOWING THE FEMORAL HEAD IS SUPEROLATERALLY POSITIONED IN RELATION TO THE CUP. VISUAL INSPECTION OF THE LINER EXHIBITS DAMAGE TO THE CONSTRAINING TABS AND RIM FEATURE. THERE ARE NICKS AND GOUGES NOTED ON THE RING AND ALSO ONE OF THE TWO CONSTRAINING TABS SHOWS DAMAGE. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MODEL #: 00889024126084. CONCOMITANT MEDICAL PRODUCTS: ITEM#: 00877702803, BIOLX OPT HD/ADPT 12/14 28+3.5, LOT#: 2955283; ITEM#: 00620105400, REPLACEMENT LOCKING RING FOR USE WITH 54 MM SHELL, LOT#: 63484424. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORT: LINER: 0001822565-2019-00353.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP PROCEDURE AND SUBSEQUENTLY 2 1/2 YEARS LATER, THE PATIENT WAS REVISED DUE TO DISLOCATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87601 CONSTRAINED LINER WITH CONSTRAINING RING HIP PROSTHESIS KWZ ZIMMER BIOMET, INC. N/A 63157006

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| R