FDA Adverse Event Malfunction Summary report: N

YC-1800

MDR report key: 8291219 · Received January 30, 2019

Report

Report Number
0002936921-2019-00003
Event Type
Malfunction
Date Received
January 30, 2019
Date of Event
January 25, 2019
Report Date
January 30, 2019
Manufacturer
NIDEK CO., LTD
Product Code
LXS
PMA / PMN Number
K893987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY NIDEK SERVICE ENGINEER (SE) ON 01/25/2019. DEVICE WAS RETURNED TO NIDEK INC. FOR SERVICE SINCE THE CUSTOMER REPORTED OF THE LASER NOT FIRING CORRECTLY. SE TESTED THE DEVICE FOR PROPER OPERATION. THE LASER WOULDN'T FIRE CONSTANTLY. EVERY TIME THE YAG LASER BEAM IS EMITTED, AN ELECTRICAL BEEP SOUNDS AND PLASMA DISCHARGE OCCURS. AN AIR PLASMA DISCHARGE OCCURS IN 8 OUT OF 10 FIRINGS AT A STANDARD OF 5 MJ OR LESS. SE OBSERVED THE LASER WOULD NOT FIRE CONSTANTLY AND PLASMA DISCHARGE OCCURRED INTERMITTENTLY. SE FOUND THAT THE ACTUAL ENERGY OUTPUT WAS HIGH; 13.5MJ. CALIBRATION WAS OUT OF SPECIFICATIONS FOR YAG LASER BEAM. HENCE, SE CALIBRATED THE SYSTEM FOR YAG LASER BEAM MAXIMUM ENERGY OUTPUT AS DESCRIBED IN THE SERVICE MANUAL SECTION 7.6.1 MAXIMUM ENERGY OUTPUT. THEN LASER OUTPUT WAS VERIFIED TO BE WITHIN SPECIFICATIONS: 0.2MJ-11.5MJ. THE CUSTOMER REPORTED ISSUE WAS DUE TO THE DEVICE OUT OF CALIBRATION. DEVICE WAS TESTED AND VERIFIED FOR PROPER OPERATION. ADDITIONALLY, THE CHIN-REST GRIP WAS BROKEN AND MISSING TWO CHIN-REST PINS. HOWEVER, THE PARTS WERE NOT REPLACED BASED ON CUSTOMER'S DECISION. THE DEVICE WAS LAST SERVICED IN 9 SEPTEMBER 2018 FOR ROUTINE PREVENTIVE MAINTENANCE. THE PRESET DATA AND OUTPUT DISPLAY CALIBRATION WAS PERFORMED. THE ENERGIES WERE WITHIN SPECIFICATION: 1.0MJ-10.5 MJ. PER OPERATORS MANUAL IT IS RECOMMENDED TO PERFORM CALIBRATION OF POWER OUTPUT OF THE YAG LASER ONCE A YEAR. REFERENCE: 2.4 AFTER USE, MAINTENANCE, AND CHECKS; CAUTION, "TO ENSURE THE CONTINUED SAFE USE OF THE DEVICE, THE MANAGER OF THE DEVICE MUST MAKE SURE THAT MAINTENANCE, PREVENTIVE INSPECTION, AND LASER OUTPUT CALIBRATION ARE PERFORMED AT LEAST ONCE A YEAR."

Description of Event or Problem · 1

NIDEK INC. RECEIVED A COMPLAINT FROM CUSTOMER ON (B)(6) 2018 REPORTING THAT DURING THE USE OF YC-1800 SN: (B)(4); EVERY 2ND TO 3RD TIME THE LASER IS FIRED, IT SOUNDS LIKE IT'S FIRING, BUT NOT FIRING. HIGH ENERGY WAS IDENTIFIED DURING THE DEVICE EVALUATION BY NIDEK INC. SERVICE ENGINEER (SE) ON 01/25/2019. NO INJURY WAS REPORTED AT THAT TIME AND HENCE NO PATIENT INFORMATION IS AVAILABLE. NIDEK INC. CONSIDERS HIGH ENERGY ISSUE A REPORTABLE EVENT AS THE DEVICE HAS MALFUNCTIONED AND HAS A POTENTIAL TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF THE ISSUE TO RECUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
83542 YC-1800 ND: YAG LASER LXS NIDEK CO., LTD YC-1800

Patients

Seq Age Sex Outcome Treatment
1