MENTOR SILTEX ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2019-08188
- Event Type
- Injury
- Date Received
- January 30, 2019
- Date of Event
- January 8, 2019
- Report Date
- January 8, 2019
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001799
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
ON 02/08/2019, THE MENTOR PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON (B)(6) 2019, A MANUFACTURING RECORD EVALUATION OF LOT NUMBER 171792 WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. ON (B)(6) 2019, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. UPON RECEIPT BY MENTOR, THE DEVICE CONTAINED CLEAR FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE AND WHITE MATERIAL WAS OBSERVED ON THE SHELL SURFACE. DURING EVALUATION THE DEVICE A RENT WITHIN SILTEX CRACKING WAS OBSERVED ON THE ANTERIOR ASPECT, MEASURING APPROXIMATELY 0.1 CM, NO OTHER ANOMALIES WERE OBSERVED. DEFLATION COMPLAINT WAS CONFIRMED. MENTOR PRODUCTS ARE 100% VISUALLY INSPECTED PRIOR TO RELEASE IN ADDITION TO THOROUGH IN-PROCESS TESTING DURING SEVERAL STAGES OF THE MANUFACTURING PROCESS MENTOR PRODUCT ANALYSIS LAB CONCLUDED THAT THE RENT AND CREASE OCCURRED SOMETIME SUBSEQUENT TO THE REMOVAL OF THE DEVICE FROM ITS PROTECTIVE PACKAGING. THE COMPLAINT WAS CONFIRMED SINCE A RENT WITHIN SILTEX CRACKING WAS FOUND ON THE DEVICE. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES CAUSED BY ACUTE FOLDS (I.E., SEVERE ¿V¿ OR TIGHT FOLDS) IN THE ROOM TEMPERATURE VULCANIZATION (RTV) SHELL OF SILTEX BREAST IMPLANTS. AT THIS TIME IS NOT POSSIBLE TO DETERMINE THE ORIGIN OF THE WHITE AND YELLOW MATERIALS. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET; HOWEVER, DEFLATION TRENDS FOR MENTOR SALINE-FILLED DEVICES WILL CONTINUE TO BE MONITORED BY QUALITY ASSURANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 325CC SALINE BREAST PROSTHESIS, CATALOG #3542650, LOT #171792. (B)(4).
IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR SILTEX ROUND MODERATE PROFILE 325CC SALINE BREAST PROSTHESIS THAT DEFLATED AFTER IMPLANTATION. DEFLATION OF THE RIGHT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2019.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 80880 | MENTOR SILTEX ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 171792 | 00081317001799 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Required Intervention | SEE H10 |