FDA Adverse Event Malfunction Summary report: N

AIRO COMPUTED TOMOGRAPHY (CT) X-RAY SYSTEM

MDR report key: 8287423 · Received January 29, 2019

Report

Report Number
3010151377-2018-00004
Event Type
Malfunction
Date Received
January 29, 2019
Date of Event
November 8, 2017
Report Date
August 31, 2018
Manufacturer
MOBIUS IMAGING, LLC
Product Code
JAK
PMA / PMN Number
K180393
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN INTERMITTENT SIGNAL FROM ONE OF THE SYSTEM DAS BOARDS WAS CAUSING PIXELS IN THE CT IMAGE TO BECOME DARK, WHICH CAUSED AN IMAGE ARTIFACT IN TWO SLICES OF THE RECONSTRUCTION OF THE CT IMAGE. THE ARTIFACTS WERE IMMEDIATELY IDENTIFIABLE TO THE OPERATOR AS A SYSTEM ISSUE/MALFUNCTION RATHER THAN AN ACTUAL PATIENT ANATOMICAL ANOMALY. THIS CASE WAS INITIALLY DETERMINED TO BE NON-REPORTABLE BY THE MANUFACTURER BECAUSE THERE WAS NO REPORT OF INJURIES AND THE RISK OF HARM TO PATIENTS IF THIS WAS TO REOCCUR WAS DETERMINED TO BE REMOTE / NOT LIKELY TO CONTRIBUTE TO A DEATH OR SERIOUS INJURY. THIS CASE IS BEING REPORTED AT THIS TIME BECAUSE DURING A FDA INSPECTION IN APRIL 2018 THIS CASE WAS SPECIFIED IN A (B)(4) AS BEING A REPORTABLE EVENT.

Description of Event or Problem · 1

IMAGE ARTIFACT IDENTIFIED BY USER IN A POST-OP PATIENT SCAN. NO ADDITIONAL SCANS OF THE PATIENT WERE REQUIRED, AND NO INJURIES REPORTED.