FDA Adverse Event Malfunction Summary report: N

BD¿¿ ORAL SYRINGE WITH TIP CAP

MDR report key: 8287380 · Received January 29, 2019

Report

Report Number
1213809-2019-00128
Event Type
Malfunction
Date Received
January 29, 2019
Date of Event
January 9, 2019
Report Date
February 15, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
KYW
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: THREE PHOTOS WERE RECEIVED AND EVALUATED. THE PHOTOS DEPICTED THREE AMBER ORAL 3ML SYRINGES WITH PLUNGER RODS PULLED UP. SAMPLES REPORTED TO BE FROM BATCH #8236737 (B)(4). ALL 3 PLUNGER RODS HAD ONE OF THEIR RIBS CRACKED NEAR THE THUMB REST. THE DAMAGE OBSERVED IS REJECTABLE PER PRODUCT SPECIFICATION. NO SAMPLES HAVE BEEN RECEIVED FOR THE OTHER 3 BATCHES. A COMPLAINT HISTORY CHECK WAS PERFORMED FOR ALL OF THE BATCHES. NO COMPLAINTS FOR ANY RELATED DEFECTS HAVE BEEN RECEIVED FROM OTHER CUSTOMERS FOR ANY OF THE BATCHES TO DATE. POTENTIAL ROOT CAUSE FOR THE DAMAGED PLUNGER ROD DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE OF THE BD¿ ORAL SYRINGE WITH TIP CAP PLUNGER RODS WERE BROKEN.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8236737. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-08-24. MEDICAL DEVICE LOT #: 8263592. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. MEDICAL DEVICE LOT #: 8263594. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. MEDICAL DEVICE LOT #: 8263595. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE OF THE BD¿ ORAL SYRINGE WITH TIP CAP PLUNGER RODS WERE BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
79207 BD¿¿ ORAL SYRINGE WITH TIP CAP LIQUID MEDICATION DISPENSER KYW BECTON DICKINSON MEDICAL SYSTEMS SEE H.10

Patients

Seq Age Sex Outcome Treatment
1 Other