BD¿¿ ORAL SYRINGE WITH TIP CAP
Report
- Report Number
- 1213809-2019-00128
- Event Type
- Malfunction
- Date Received
- January 29, 2019
- Date of Event
- January 9, 2019
- Report Date
- February 15, 2019
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- KYW
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: THREE PHOTOS WERE RECEIVED AND EVALUATED. THE PHOTOS DEPICTED THREE AMBER ORAL 3ML SYRINGES WITH PLUNGER RODS PULLED UP. SAMPLES REPORTED TO BE FROM BATCH #8236737 (B)(4). ALL 3 PLUNGER RODS HAD ONE OF THEIR RIBS CRACKED NEAR THE THUMB REST. THE DAMAGE OBSERVED IS REJECTABLE PER PRODUCT SPECIFICATION. NO SAMPLES HAVE BEEN RECEIVED FOR THE OTHER 3 BATCHES. A COMPLAINT HISTORY CHECK WAS PERFORMED FOR ALL OF THE BATCHES. NO COMPLAINTS FOR ANY RELATED DEFECTS HAVE BEEN RECEIVED FROM OTHER CUSTOMERS FOR ANY OF THE BATCHES TO DATE. POTENTIAL ROOT CAUSE FOR THE DAMAGED PLUNGER ROD DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED.
IT WAS REPORTED THAT BEFORE USE OF THE BD¿ ORAL SYRINGE WITH TIP CAP PLUNGER RODS WERE BROKEN.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8236737. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-08-24. MEDICAL DEVICE LOT #: 8263592. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. MEDICAL DEVICE LOT #: 8263594. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. MEDICAL DEVICE LOT #: 8263595. MEDICAL DEVICE EXPIRATION DATE: N/A. DEVICE MANUFACTURE DATE: 2018-09-20. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BEFORE USE OF THE BD¿ ORAL SYRINGE WITH TIP CAP PLUNGER RODS WERE BROKEN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 79207 | BD¿¿ ORAL SYRINGE WITH TIP CAP | LIQUID MEDICATION DISPENSER | KYW | BECTON DICKINSON MEDICAL SYSTEMS | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |