FDA Adverse Event Malfunction Summary report: N

METRIX AIR

MDR report key: 8282467 · Received January 28, 2019

Report

Report Number
1000113657-2019-00064
Event Type
Malfunction
Date Received
January 28, 2019
Date of Event
January 3, 2019
Report Date
January 28, 2019
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007720
PMA / PMN Number
K150052
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. RETURNED TEST STRIPS EVALUATED WITH DEFECT FOUND. QC INVESTIGATION ON 1/10/2019 RETURNED TEST STRIPS PRODUCE HIGH READINGS. R&D FINAL ROOT CAUSE: MOST LIKELY UNDERLYING ROOT CAUSE OF HIGH READINGS ARE DUE TO IMPROPER STORAGE: VIAL LEFT OPEN FOR AN EXTENDED PERIOD OF TIME TEST STRIP UDI# (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON (B)(6) 2019, IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION OR SYMPTOMS SINCE THE LAST CALL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS 150, 186, 252, 284 AND 180 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 120 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2019, A BACK-TO-BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. CUSTOMER ALSO STATED THAT HE REMOVED THE TEST STRIPS FROM THE VIAL AND PLACED IN THE CARRYING CASE. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 02/29/2020, AND OPEN VIAL DATE IS ONE WEEK. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75863 METRIX AIR BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. METRIX AIR MV2909 00021292007720

Patients

Seq Age Sex Outcome Treatment
1 0 YR