FDA Adverse Event Malfunction Summary report: N

NEEDLE 21X1 RB

MDR report key: 8280008 · Received January 25, 2019

Report

Report Number
1911916-2019-00118
Event Type
Malfunction
Date Received
January 25, 2019
Date of Event
January 7, 2019
Report Date
March 1, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
00382903051656
PMA / PMN Number
K021475
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: TEN REPRESENTATIVE SAMPLES WERE RECEIVED FOR EVALUATION. THEY WERE VISUALLY INSPECTED. THE SEALING OF PACKAGING IS GOOD AND ALL THE PLASTIC SHIELD NEEDLES ARE ASSEMBLED. THEY ALL WERE TESTED FOR SHIELD PULL TEST. THE VALUES WERE BETWEEN 2.2LBS TO 2.5LBS; THE SPECIFICATION IS 0.5LBS TO 5.5LBS, THEREFORE FAILURE MODE IS NOT VERIFIED. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF NEEDLE 21X1 RB HAD THE NEEDLE CAPS FALL OFF DURING USE.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(6). THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8090568. MEDICAL DEVICE EXPIRATION DATE: 2023-05-31. DEVICE MANUFACTURE DATE: 2018-03-31. MEDICAL DEVICE LOT #: 8204603. MEDICAL DEVICE EXPIRATION DATE: 2023-09-30. DEVICE MANUFACTURE DATE: 2018-07-23. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF NEEDLE 21X1 RB HAD THE NEEDLE CAPS FALL OFF DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69905 NEEDLE 21X1 RB HYPODERMIC SINGLE LUMEN NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL SEE H.10 00382903051656

Patients

Seq Age Sex Outcome Treatment
1 Other