FDA Adverse Event Malfunction Summary report: N

ULTRASAFE X100L PNG CLEAR SDZ

MDR report key: 8279785 · Received January 25, 2019

Report

Report Number
3009081593-2019-00101
Event Type
Malfunction
Date Received
January 25, 2019
Date of Event
January 18, 2019
Report Date
January 24, 2019
Manufacturer
BECTON DICKINSON HUNGARY KFT (BD)
Product Code
MEG
UDI-DI
00382908607049
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. PMA / 510(K)#: K011369, K122558. INVESTIGATION SUMMARY: THE CUSTOMER ISSUED A COMPLAINT FOR A CAN'T INJECT PROBLEM DETECTED BY END USER. NEITHER PHOTO NOR SAMPLE WAS PROVIDED TO BD MEDICAL ¿ PHARMACEUTICAL SYSTEM (BDM-PS) FOR ANALYSIS. BDM-PS PERFORMED A BATCH HISTORY RECORD¿S REVIEW (BHR) INCLUDING A REVIEW OF ALL DATA COLLECTED DURING IN PROCESS AND QUALITY INSPECTIONS. THE BATCHES INVOLVED IN THIS COMPLAINT MEET ALL ACCEPTABLE QUALITY LEVELS (AQL¿S), WERE MANUFACTURED AND RELEASED ACCORDING TO APPLICABLE PROCEDURES AND SPECIFICATIONS. IT SHOULD BE NOTED THAT TESTS PERFORMED BY BD (B)(4) DURING PRODUCTION CONTROLS ARE RELATED TO THE ASSEMBLED DEVICE, WITHOUT USING SYRINGE AND PLUNGER ROD. TESTING OF COMBINATION PRODUCT IS OUT OF SCOPE FOR SAFETY DEVICE MANUFACTURING. BASED ON INVESTIGATION CONCLUSION, BDM-PS WAS NOT ABLE TO CONFIRM THE SYMPTOM PERCEIVED BY CUSTOMER OR CORRELATE THIS SYMPTOM WITH A POTENTIAL CAUSE LINKED TO BD PROCESS. INVESTIGATION CONCLUSION: UNCONFIRMED, NO ISSUE OBSERVED. ROOT CAUSE DESCRIPTION:" BASED ON INVESTIGATION CONCLUSION, BDM-PS WAS NOT ABLE TO CONFIRM THE SYMPTOM PERCEIVED BY CUSTOMER OR CORRELATE THIS SYMPTOM WITH A POTENTIAL CAUSE LINKED TO BD PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER INJECTION THE SAFETY DEVICE ON A ULTRASAFE X100L PNG CLEAR SDZ FAILED TO ACTIVATE AND RETRACT THE NEEDLE FROM THE INJECTION SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
69856 ULTRASAFE X100L PNG CLEAR SDZ PISTON SYRINGE MEG BECTON DICKINSON HUNGARY KFT (BD) 7300818 00382908607049

Patients

Seq Age Sex Outcome Treatment
1 Other