MILAGRO ADVANCE SCREW 9X30MM
Report
- Report Number
- 1221934-2018-53250
- Event Type
- Malfunction
- Date Received
- January 23, 2019
- Date of Event
- January 30, 2017
- Report Date
- January 30, 2017
- Manufacturer
- DEPUY MITEK LLC US
- Product Code
- HWC
- UDI-DI
- 10886705022274
- PMA / PMN Number
- K123362
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY. CORRECTION: H4: DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE WAS INADVERTENTLY MISSED IN THE INITIAL REPORT AND HAS BEEN UPDATED AS 8/18/2015. THIS REPORT IS BEING FILED FROM THE ETQ COMPLAINT MANAGEMENT SYSTEM AS REQUIRED UNDER MITEK'S CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) TO FILE USA FDA MDR MISSED MALFUNCTIONS.
UDI: (B)(4). THE COMPLAINT DEVICE IS NOT BEING RETURNED, THEREFORE UNAVAILABLE FOR A PHYSICAL EVALUATION. A REVIEW INTO THE DEPUY SYNTHES MITEK COMPLAINTS SYSTEM REVEALED NO OTHER COMPLAINTS FOR THIS LOT OF DEVICES THAT WERE RELEASED TO DISTRIBUTION. WE CANNOT DISCERN A ROOT CAUSE FOR THE REPORTED FAILURE MODE. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED AND NO FURTHER ACTION IS WARRANTED. THIS FILE WILL REMAIN RECEPTIVE TO ANY POTENTIAL FORTHCOMING INFORMATION RECEIVED THAT IS PERTINENT AND GERMANE TO THIS ISSUE. A NON-CONFORMANCE SEARCH WAS PERFORMED FOR THIS PRODUCT CODE 231822-LOT #3860688 COMBINATION AND NO NON-CONFORMANCES WERE IDENTIFIED. NO FURTHER INFORMATION REGARDING THE TECHNIQUE OR INSTRUMENTS USED HAS BEEN PROVIDED TO DETERMINE A ROOT CAUSE FOR THIS FAILURE. IF ANY ADDITIONAL INFORMATION IS OBTAINED, THIS COMPLAINT WILL BE RE-OPENED TO CAPTURE THAT INFORMATION. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. THIS REPORT IS BEING FILED FROM THE (B)(4) AS REQUIRED UNDER MITEK'S CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) TO FILE USA FDA MDR MISSED MALFUNCTIONS.
IT WAS REPORTED BY THE SALES REP THAT THE 9X30 MILAGRO TAPERED SCREW DEVICE BROKE WHEN BEING INSERTED INTO THE PATIENT DURING AN ANTERIOR CRUCIATE LIGAMENT ACL SURGICAL PROCEDURE. THE CASE WAS COMPLETED USING ANOTHER SCREW IN THE SAME BONE HOLE. THERE WERE NO PATIENT CONSEQUENCES BUT A 5 MINUTE DELAY WAS REPORTED. THE SALES REP STATED THAT THE DEVICE WAS DISCARDED BY THE CUSTOMER AND NOT AVAILABLE FOR RETURN. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 63990 | MILAGRO ADVANCE SCREW 9X30MM | ORTHOPAEDIC BONE SCREW, BIOABSORBABLE | HWC | DEPUY MITEK LLC US | 3860688 | 10886705022274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |