FDA Adverse Event Malfunction Summary report: N

SERVO-I

MDR report key: 8270596 · Received January 23, 2019

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
8010042-2019-00049
Event Type
Malfunction
Date Received
January 23, 2019
Date of Event
January 9, 2019
Report Date
March 11, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K123149
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERON: (B)(4). THE VENTILATOR WAS INVESTIGATED ON SITE AND THE O2 GAS MODULE WAS REPLACED AND RETURNED FOR INVESTIGATION. DURING SIMULATED USE TESTING OF THE RETURNED PART IN A VENTILATOR SYSTEM, IT WAS NOT ABLE TO REPRODUCE THE REPORTED O2 CONCENTRATION LOW. UNIT TESTING OF THE GAS MODULE WE WERE ALSO PERFORMED, BUT NO DEVIATIONS WERE FOUND. THERE WAS 2 ALARMS CONFIRMED FOR O2 CONCENTRATION LOW, AT JANUARY 5 21:18 AND 21:25, THEREFORE THE DATE OF EVENT WAS DETERMINED AS (B)(6) 2018. THE INVESTIGATION FINDINGS DO NOT LEAD TO A CLEAR CONCLUSION ABOUT THE CAUSE OF THE REPORTED EVENTS.

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERON: (B)(4).

Description of Event or Problem · 0

MANUFACTURER REF. #: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING PATIENT TREATMENT, THE VENTILATOR GENERATED ALARMS FOR LOW O2 CONCENTRATION. THERE WAS NO PATIENT HARM. (B)(4).