MENTOR SILTEX CONTOUR PROFILE SPECTRUM

MDR report key: 8264446 · Received January 21, 2019

Report

Report Number
1645337-2019-08017
Event Type
Injury
Date Received
January 21, 2019
Date of Event
December 31, 2018
Report Date
January 2, 2019
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317002178
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 2/20/19, DEVICE EVALUATION AND INVESTIGATION FINDINGS: UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. MENTOR PRODUCT ANALYSIS LAB DISCOVERED A RENT MEASURING APPROXIMATELY 0.4 CM WITHIN A LINE OF SHELL WEAR, SUGGESTING IN-VIVO FOLDING OR CREASING OF THE DEVICE. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. A MANUFACTURING RECORD EVALUATION (MRE) OF LOT NUMBER 5934738 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BECAUSE MENTOR PRODUCT ANALYSIS LAB WAS UNABLE TO CONFIRM ANY UNUSUAL FAILURE MODE, NO CORRECTIVE ACTION WILL BE TAKEN AT THIS TIME. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST RECONSTRUCTION PRIMARY WITH A MENTOR SILTEX CONTOUR PROFILE SPECTRUM 650CC AND EXPERIENCED DEFLATION OF THE LEFT BREAST TISSUE EXPANDER. AS A RESULT, THE PATIENT HAD UNDERGONE EXPLANTATION ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
57459 MENTOR SILTEX CONTOUR PROFILE SPECTRUM PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5934738 00081317002178

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention