VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 60/28
Report
- Report Number
- 3005180920-2018-01109
- Event Type
- Injury
- Date Received
- January 18, 2019
- Date of Event
- December 20, 2018
- Report Date
- January 18, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- MEH
- UDI-DI
- 07630030807473
- PMA / PMN Number
- K092265
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 17 JANUARY 2019. LOT 152106: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17 JUNE 2015 . EXPIRATION DATE: 2020-04-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR ON JANUARY 17, 2019. 3 YEARS AFTER PRIMARY CEMENTLESS DM THR AN INTRAPROSTHETIC DISLOCATION OCCURRED. THE CERAMIC HEAD, REPORTEDLY MANUFACTURED BY A DIFFERENT COMPANY, DISLOCATED FROM THE PE LINER. THERE ARE NO INDICATIONS AS TO POSSIBLE CAUSES FOR THIS EVENT, SUCH AS TRAUMAS OR EXTREME MOVEMENTS. WE COULD NOT MEASURE EITHER THE HEAD OF THE PE INSERT AND VERIFY IF A TOLERANCE MISMATCH OCCURRED, BUT IT IS UNLIKELY GIVEN THE SIGNIFICANT TIME AFTER PRIMARY IMPLANTATION. NO DEFINITE REASON FOR THIS EVENT COULD BE ASCERTAINED WITH THE INFORMATION AT HAND. PRELIMINARY INVESTIGATION PERFORMED BY R&D PRODUCT MANAGER ON JANUARY 14, 2019 BY R&D PRODUCT MANAGER: PRELIMINARY VISUAL INVESTIGATION ON EXPLANTED COMPONENT SHOWS SIGNS OF FRICTION BETWEEN FEMORAL HEAD AND METAL CUP, RESULTING IN METAL TRANSFER ON FEMORAL HEAD AND WEAR OF THE ACETABULAR COMPONENT'S EDGE AND ARTICULATING SURFACE. SCRATCHES SIGNS ON THE PE LINER EDGE DUE TO EXTRACTION AND NON-COHERENT FUNCTION AFTER DISLOCATION.
THE PATIENT CAME IN, 3 YEARS AND 1 MONTH AFTER PRIMARY SURGERY, DUE TO INTRA-PROSTHETIC DISLOCATION OF THE HEAD FROM THE LINER. THE SURGEON REVISED THE MEDACTA CUP, MEDACTA LINER AND ANOTHER COMPANY'S HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 54513 | VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 60/28 | DOUBLE MOBILITY HC LINER | MEH | MEDACTA INTERNATIONAL SA | 152106 | 07630030807473 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |