FDA Adverse Event
Malfunction
Summary report: N
FC 500 FLOW CYTOMETER
MDR report key: 8259184
·
Received January 17, 2019
Report
- Report Number
- 1061932-2019-01210
- Event Type
- Malfunction
- Date Received
- January 17, 2019
- Date of Event
- September 18, 2018
- Report Date
- January 17, 2019
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- UDI-DI
- 15099590483531
- PMA / PMN Number
- K071681
- Removal / Correction Number
- 2050012-0108/2018-001C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE ENGINEER (FSE) WAS ON SITE AND CONFIRMED WHAT THE CUSTOMER REPORTED. TO RESOLVE THE ISSUE THE FSE REPLACED THE TARPON AMP BOARD. BEC IS FILING AN MDR FOR THIS EVENT BASED ON THE FDA CLASSIFICATION OF THE 08 JAN 2018 URGENT MEDICAL DEVICE RECALL AS A CLASS I RECALL ON 20 NOV 2018 (RECALL NUMBER Z-0471-2019 FOR FC 500; RECALL NUMBER Z-0472-2019 FOR EPICS XL/XLMCL). BEC INTERNAL IDENTIFIER ¿ (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED NO SIGNAL DETECTED AT FL2 ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51277 | FC 500 FLOW CYTOMETER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | FC500, RECOND TN, MCL UPS FLOW CYTOMETER 100-240V, 50/60HZ | 15099590483531 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |